Hakenberg O W, Ryall R L, Langlois S L, Marshall V R
J Urol. 1983 Aug;130(2):249-51. doi: 10.1016/s0022-5347(17)51087-3.
Real-time ultrasonography was used to measure bladder volumes. Volumes were calculated as the product of 3 internal bladder diameters (height, width and depth). The true bladder volume was obtained from the voided volume or by catheterization. There was a good correlation between calculated and true volumes provided a correction factor of 0.6 was applied. It also was found that a simpler method using only 2 diameters and a correction factor of 0.15 was nearly as reliable. The accuracy of this quantitative method was shown to be limited to an average error of plus or minus 25 per cent for bladder volumes between 100 and 500 ml. Smaller volumes could be assessed only qualitatively. False negative readings were common for bladder volumes less than 50 ml.
采用实时超声测量膀胱容量。容量计算为膀胱三个内部直径(高度、宽度和深度)的乘积。通过排尿量或导尿获得真实膀胱容量。如果应用0.6的校正因子,计算容量与真实容量之间具有良好的相关性。还发现,使用仅两个直径和0.15的校正因子的更简单方法几乎同样可靠。对于100至500毫升之间的膀胱容量,该定量方法的准确性显示平均误差限制在正负25%。较小容量只能进行定性评估。对于小于50毫升的膀胱容量,假阴性读数很常见。