A 90-year-old woman received a dual chamber pacemaker (PM) for a sick sinus syndrome. The PM was programmed with SafeR AAI-DD pacing mode at 60 bpm. During a standard follow up, some memorized electrograms (EGMs) were found in SafeR diagnostics, with atrial pacing (Ap) not followed by any ventricular sensing/pacing event, due to simultaneous junctional activity falling into ventricular blanking period during Ap and, for this reason, unsensed by the PM. Blanking periods can affect PM functioning if not revealed and adjusted.
一名90岁女性因病态窦房结综合征植入了双腔起搏器(PM)。该起搏器采用SafeR AAI-DD起搏模式,设置频率为60次/分钟。在一次标准随访中,SafeR诊断功能发现了一些存储的心电图(EGM),心房起搏(Ap)后未出现任何心室感知/起搏事件,原因是在心房起搏期间,同时出现的交界性活动落入心室空白期,因此起搏器未感知到。如果未发现并调整空白期,可能会影响起搏器功能。