Department of Urology, Beijing Chaoyang Hospital, Capital Medical University, Institute of Urology, Capital Medical University, Beijing, China (mainland).
Department of Urology, The First Central Hospital of Mongolia, Ulaanbaatar, Mongolia.
Ann Transplant. 2020 Nov 27;25:e928773. doi: 10.12659/AOT.928773.
BACKGROUND This study investigated the safety and efficacy of a new cooling device for use in minimal-incision kidney transplantation (MIKT). MATERIAL AND METHODS From June 2016 to December 2021, 9 patients underwent MIKT surgery in our hospitals using the new cooling device to maintain hypothermia. We recorded and analyzed information on the etiology, comorbid status, ongoing renal replacement assessment, BMI, HLA mismatching sites of donors and recipients, and perioperative and postoperative clinical data for recipients. RESULTS Kidney transplantation was successfully performed in all patients. The kidney surface temperature measurement results showed that the intraoperative renal anterior and posterior surface temperatures were stable at approximately 3.8±1.2°C and 5.2±1.3°C, respectively, during ice-water circulation. The mean operation time was 112±15 min, the artery anastomosis time was 16±6.0 min, and the vein anastomosis time was 14±4.5 min. All recipients recovered uneventfully. The patients were followed up for 6-30 months. Urinary and vascular complications were not found in any recipients. CONCLUSIONS The new cooling device can facilitate MIKT. It is safe and feasible to carry out MIKT using the new cooling device, which can reduce surgical trauma and improve the quality of vascular anastomosis with satisfactory cosmetic results.
本研究旨在探讨一种新的冷却装置在微创肾移植(MIKT)中的安全性和有效性。
自 2016 年 6 月至 2021 年 12 月,我院采用新的冷却装置对 9 例患者实施 MIKT 手术,以维持低温。我们记录并分析了患者的病因、合并症状况、持续的肾脏替代评估、BMI、供体和受者 HLA 错配部位以及受者围手术期和术后的临床数据。
所有患者均成功进行了肾移植。术中肾前、后表面温度测量结果显示,在冰水循环过程中,肾前、后表面温度分别稳定在 3.8±1.2°C 和 5.2±1.3°C 左右。手术时间平均为 112±15min,动脉吻合时间为 16±6.0min,静脉吻合时间为 14±4.5min。所有受者均恢复顺利。患者随访 6-30 个月,未发现受者出现尿及血管并发症。
新型冷却装置可促进 MIKT。使用新型冷却装置开展 MIKT 安全可行,可减少手术创伤,改善血管吻合质量,美容效果满意。