Evaluation of the Datascope ACCUSAT pulse oximeter in healthy adults.

作者信息

Mendelson Y, Kent J C, Shahnarian A, Welch G W, Giasi R M

机构信息

Biomedical Engineering Program, Worcester Polytechnic Institute, MA 01609.

出版信息

J Clin Monit. 1988 Jan;4(1):59-63. doi: 10.1007/BF01618108.

Abstract

Arterial hemoglobin oxygen saturation measured with the Datascope ACCUSAT pulse oximeter was compared with simultaneous arterial hemoglobin oxygen saturation measurements in healthy adult volunteers. One hundred thirty-five arterial blood samples ranging in saturation from 63 to 100% were obtained from 15 adults, aged 20 to 43 years. These subjects had different skin pigmentation, hematocrit, and smoking habits. Steady-state hypoxia was achieved by varying the inspired oxygen concentration between 10 and 21%. Readings from the Datascope ACCUSAT pulse oximeter and the Hewlett-Packard 47201A ear oximeter were compared with arterial blood samples analyzed by the Instrumentation Laboratories IL 282 in vitro CO-Oximeter. The equation for the best fitted linear regression line between the ACCUSAT pulse oximeter and the reference IL 282 CO-Oximeter was: ACCUSAT = 1.08(IL) - 6.86. The linear regression analysis revealed a high degree of correlation (r = 0.99), and a small standard error of the estimate (SEE = 1.29%). Simultaneous comparisons between arterial hemoglobin oxygen saturation measured with the ACCUSAT pulse oximeter and the Hewlett-Packard ear oximeter also showed a close correlation (r = 0.99, SEE = 1.47%). A similar comparison between the ACCUSAT and the Ohmeda 3700 pulse oximeter revealed good correlation (r = 0.99, SEE = 1.72%). We found that the ACCUSAT pulse oximeter is an accurate instrument for measuring arterial hemoglobin oxygen saturation noninvasively in the range between 60 and 100%.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验