A prior publication introduced the Strome-Blitzer balloon's ability to obtain circumferential esophageal cytologic sampling. This GLP study was requisite for FDA approval to determine if equivalent cell capture and cellularity was observed with the balloon compared to surface sampling brushes and to determine the balloon's usability for naive otolaryngologists.
Three naïve users tested the Hobbs brush and Strome-Blitzer balloon on 4 Yorkshire swine. Four anatomical sites were sampled, beginning distally and ending proximally. In 2 animals, the balloon was used first distally and in the remaining 2, 4 new Hobbs brushes were used distally first. Moving proximally, the balloon and brushes were sequentially alternated. In follow-the-leader fashion, the balloon was introduced trans-orally followed by an endoscope to the desired site. The balloon was inflated exposing the abrasive strips to contact the esophageal mucosa. Moving the balloon 1 to 2 cm superiorly and inferiorly effected circumferential cell capture. The balloon was collapsed and removed, preserving the cellularity. The Hobbs brush was passed through the scope's channel. Four brushes, 1 per quadrant, obtained the samples at an anatomical site. The balloon was rated as pass/fail on the following: delivery, kinking, usability, and malfunction. A blinded veterinary pathologist evaluated the cytology.
There was no device malfunction, mucosal trauma, or difficulty with device use. Balloon cytologic samples were comparable in cellularity and quality to the brush.
A single balloon sampling was comparable to 4 brushes in capturing diagnostically relevant cellular volumes and architecture. Naïve users easily performed the procedures after reading the guidelines.
先前的出版物介绍了 Strome-Blitzer 球囊获取食管细胞学全周采样的能力。这项 GLP 研究是 FDA 批准所必需的,旨在确定与表面采样刷相比,球囊是否能观察到等效的细胞捕获和细胞密度,并确定球囊对新手耳鼻喉科医生的可用性。
3 名新手用户在 4 只约克夏猪上测试了 Hobbs 刷和 Strome-Blitzer 球囊。在 4 个解剖部位进行采样,从远端开始,到近端结束。在 2 只动物中,首先在远端使用球囊,在其余 2 只动物中,首先在远端使用 4 个新的 Hobbs 刷。向近端移动时,球囊和刷子依次交替使用。以追随领导者的方式,球囊经口内引入,然后内镜到达所需部位。球囊充气,使研磨带接触食管黏膜。上下移动球囊 1 至 2 厘米可实现全周细胞捕获。球囊被压缩并取出,保留细胞密度。Hobbs 刷穿过内镜的通道。4 个刷子,每个象限 1 个,在一个解剖部位获取样本。球囊的评价标准为:输送、扭结、可用性和故障。一位盲法兽医病理学家对细胞学进行了评估。
没有设备故障、黏膜创伤或使用设备困难。球囊细胞学样本在细胞密度和质量上与刷子相当。
单次球囊采样与 4 个刷子相比,可捕获具有诊断意义的细胞体积和结构。新手用户在阅读指南后,很容易进行这些操作。
3 级。