Arch Intern Med. 1988 Mar;148(3):673-6.

PMID:3341868
Abstract

We studied 36 hypertensive men, aged 60 years or older, and compared direct intra-arterial diastolic blood pressure (BP) measurements, indirect cuff mercury sphygmomanometer measurements, and automatic infrasonic recorder (IR) measurements. We used a receiver operating characteristic curve to determine whether a difference between the cuff and IR diastolic BP could identify patients likely to have pseudo-hypertension (cuff-intra-arterial diastolic BP difference of 10 mm Hg or greater). We found that a cuff-IR difference of 4 mm Hg could identify the majority of patients with pseudo-hypertension (sensitivity, 93%; specificity, 64%; positive predictive value, 62%; negative predictive value, 93%). We concluded that the IR is an accurate, quantitative, noninvasive substitute for intra-arterial measurements and may be the preferred technique for screening and monitoring of blood pressure in elderly hypertensive patients.

摘要

我们研究了36名60岁及以上的高血压男性,比较了直接动脉内舒张压(BP)测量值、间接袖带汞血压计测量值和自动次声记录仪(IR)测量值。我们使用受试者工作特征曲线来确定袖带舒张压与IR舒张压之间的差异是否能够识别可能患有假性高血压(袖带-动脉内舒张压差异为10毫米汞柱或更大)的患者。我们发现,4毫米汞柱的袖带-IR差异能够识别大多数假性高血压患者(敏感性为93%;特异性为64%;阳性预测值为62%;阴性预测值为93%)。我们得出结论,IR是一种准确、定量、无创的动脉内测量替代方法,可能是老年高血压患者血压筛查和监测的首选技术。