The aim of this study was to compare the visual and refractive outcomes of total keratometry (TK) versus anterior keratometry (AK) measurements of the IOLMaster 700® (Carl Zeiss Meditec AG, Jena, Germany) in surgery for age-related cataract with preexisting corneal astigmatism.
Monocentric retrospective comparative study. The IOLMaster 700® biometer was used in the 2 groups: in AK mode (AK group) and in TK mode (TK group), for toric IOL (AT TORBI 709 MP) calculation with ZCALC®, Zeiss toric IOL calculator. A 2:1 matching was made between the AK and TK groups. Uncorrected distance visual acuity (UDVA), the correction index and the error in predicted residual astigmatism were analyzed 1 month postoperatively using the vector analysis by the Alpins method.
The whole cohort included 405 eyes distributed as follows after 2:1 matching: 158 eyes in the AK group and 79 eyes in the TK group. The mean UDVA was similar in both groups (0.07 ± 0.10 LogMAR; p = 0.587). No significant difference in mean absolute error in predicted residual astigmatism (0.37 ± 0.33 D versus 0.35 ± 0.26 D; p = 0.545) and in mean centroid error in predicted residual astigmatism (0.19 ± 0.49 at 3° and 0.06 ± 0.46 at 0°; p = 0.008 and 0.161 respectively for the x- and y-components) was found between the AK and TK groups.
TK of the IOLMaster 700® gives excellent refractive and visual outcomes, comparable to those obtained in AK mode, without showing its superiority for corneas with regular astigmatism.
本研究旨在比较有晶状体眼前房型人工晶状体(IOL)计算中,应用 IOLMaster700®(德国卡尔蔡司公司)的角膜整体(TK)和角膜前表面(AK)测量值的视觉和屈光结果,以评估其在伴有角膜散光的年龄相关性白内障手术中的应用。
这是一项单中心回顾性对比研究。IOLMaster700®生物测量仪用于两组:AK 模式(AK 组)和 TK 模式(TK 组),并采用 ZCALC®(蔡司散光人工晶状体计算器)计算 Toric IOL(AT TORBI709MP)。AK 组和 TK 组之间进行了 2:1 匹配。术后 1 个月,采用 Alpins 矢量分析法,分析两组的未矫正远视力(UDVA)、矫正指数和预测残余散光误差。
全队列共纳入 405 只眼,2:1 匹配后分为两组:AK 组 158 只眼,TK 组 79 只眼。两组的 UDVA 均值相似(0.07 ± 0.10 LogMAR;p = 0.587)。预测残余散光的平均绝对误差(0.37 ± 0.33 D 对 0.35 ± 0.26 D;p = 0.545)和预测残余散光的平均质心误差(x 轴和 y 轴在 3°和 0°时分别为 0.19 ± 0.49 和 0.06 ± 0.46;p = 0.008 和 0.161)在 AK 组和 TK 组之间无显著差异。
IOLMaster700®的 TK 测量值可提供出色的屈光和视觉效果,与 AK 模式相当,对于规则性散光角膜无优势。