在0.35T MR-LINAC上多个机架角度下对扩散加权磁共振成像和几何畸变的评估。

Evaluation of diffusion-weighted MRI and geometric distortion on a 0.35T MR-LINAC at multiple gantry angles.

作者信息

Lewis Benjamin, Guta Anamaria, Mackey Stacie, Gach H Michael, Mutic Sasa, Green Olga, Kim Taeho

机构信息

Departments of Radiation Oncology, Washington University School of Medicine in St. Louis, St. Louis, MO, USA.

Department of Radiation Oncology, Barnes Jewish Hospital, St. Louis, MO, USA.

出版信息

J Appl Clin Med Phys. 2021 Feb;22(2):118-125. doi: 10.1002/acm2.13135. Epub 2021 Jan 15.

Abstract

Diffusion-weighted imaging (DWI) provides a valuable diagnostic tool for tumor evaluation. Yet, it is difficult to acquire daily MRI data sets in the traditional radiotherapy clinical setting due to patient burden and limited resources. However, integrated MRI radiotherapy treatment systems facilitate daily functional MRI acquisitions like DWI during treatment exams. Before ADC values from MR-RT systems can be used clinically their reproducibility and accuracy must be quantified. This study used a NIST traceable DWI phantom to verify ADC values acquired on a 0.35 T MR-LINAC system at multiple gantry angles. A diffusion-weighted echo planar imaging sequence was used for all image acquisitions, with b-values of 0, 500, 900, 2000 s/mm for the 1.5 T and 3.0 T systems and 0, 200, 500, 800 s/mm for the 0.35 T system. Images were acquired at multiple gantry angles on the MR-LINAC system from 0° to 330° in 30° increments to assess the impact of gantry angle on geometric distortion and ADC values. CT images, and three fiducial markers were used as ground truth for geometric distortion measurements. The distance between fiducial markers increased by as much as 7.2 mm on the MR-LINAC at gantry angle 60°. ADC values of deionized water vials from the 1.5 T and 3.0 T systems were 8.30 × 10  mm /s and -0.85 × 10  mm /s off, respectively, from the expected value of 1127 × 10  mm /s. The MR-LINAC system provided an ADC value of the pure water vials that was -116.63 × 10  mm /s off from the expected value of 1127 × 10  mm /s. The MR-LINAC also showed a variation in ADC across all gantry angles of 33.72 × 10  mm /s and 20.41 × 10  mm /s for the vials with expected values of 1127 × 10  mm /s and 248 × 10  mm /s, respectively. This study showed that variation of the ADC values and geometric information on the 0.35 T MR-LINAC system was dependent on the gantry angle at acquisition.

摘要

扩散加权成像(DWI)为肿瘤评估提供了一种有价值的诊断工具。然而,在传统放疗临床环境中,由于患者负担和资源有限,难以获取每日的MRI数据集。然而,集成MRI放疗治疗系统有助于在治疗检查期间进行每日的功能MRI采集,如DWI。在将MR-RT系统的ADC值用于临床之前,必须对其再现性和准确性进行量化。本研究使用了一个可溯源至美国国家标准与技术研究院(NIST)的DWI体模,以验证在0.35 T MR-LINAC系统上多个机架角度采集的ADC值。所有图像采集均使用扩散加权回波平面成像序列,1.5 T和3.0 T系统的b值为0、500、900、2000 s/mm²,0.35 T系统的b值为0、200、500、800 s/mm²。在MR-LINAC系统上以30°增量从0°到330°的多个机架角度采集图像,以评估机架角度对几何畸变和ADC值的影响。CT图像和三个基准标记用作几何畸变测量的参考标准。在机架角度为60°时,MR-LINAC上基准标记之间的距离增加了多达7.2 mm。1.5 T和3.0 T系统中去离子水瓶的ADC值分别比预期值1127×10⁻⁶ mm²/s偏离8.30×10⁻⁶ mm²/s和 -0.85×10⁻⁶ mm²/s。MR-LINAC系统提供的纯水瓶的ADC值比预期值1127×10⁻⁶ mm²/s偏离 -116.63×10⁻⁶ mm²/s。对于预期值分别为1127×10⁻⁶ mm²/s和248×10⁻⁶ mm²/s的水瓶,MR-LINAC在所有机架角度上还显示出ADC值的变化,分别为33.72×10⁻⁶ mm²/s和20.41×10⁻⁶ mm²/s。本研究表明,0.35 T MR-LINAC系统上ADC值和几何信息的变化取决于采集时的机架角度。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/2ecc/7882099/5cfc13f6e8ca/ACM2-22-118-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验