专用乳腺正电子发射断层扫描探测器边缘图像质量的评估。

Evaluation of image quality at the detector's edge of dedicated breast positron emission tomography.

作者信息

Satoh Yoko, Motosugi Utaroh, Imai Masamichi, Omiya Yoshie, Onishi Hiroshi

机构信息

Yamanashi PET Imaging Clinic, Shimokato 3046-2, Chuo City, Yamanashi Prefecture, 409-3821, Japan.

Department of Radiology, University of Yamanashi, Chuo City, Yamanashi Prefecture, Japan.

出版信息

EJNMMI Phys. 2021 Jan 18;8(1):5. doi: 10.1186/s40658-020-00351-6.

Abstract

BACKGROUND

Using phantoms and clinical studies in prone hanging breast imaging, we assessed the image quality of a commercially available dedicated breast PET (dbPET) at the detector's edge, where mammary glands near the chest wall are located. These are compared to supine PET/CT breast images of the same clinical subjects.

METHODS

A breast phantom with four spheres (16-, 10-, 7.5-, and 5-mm diameter) was filled with F-fluorodeoxyglucose solution (sphere-to-background activity concentration ratio, 8:1). The spheres occupied five different positions from the top edge to the centre of the detector and were scanned for 5 min in each position. Reconstructed images were visually evaluated, and the contrast-to-noise ratio (CNR), contrast recovery coefficient (CRC) for all spheres, and coefficient of variation of the background (CV) were calculated. Subsequently, clinical images obtained with standard supine PET/CT and prone dbPET were retrospectively analysed. Tumour-to-background ratios (TBRs) between breast cancer near the chest wall (close to the detector's edge; peripheral group) and at other locations (non-peripheral group) were compared. The TBR of each lesion was compared between dbPET and PET/CT.

RESULTS

Closer to the detector's edge, the CNR and CRC of all spheres decreased while the CV increased in the phantom study. The disadvantages of this placement were visually confirmed. Regarding clinical images, TBR of dbPET was significantly higher than that of PET/CT in both the peripheral (12.38 ± 6.41 vs 6.73 ± 3.5, p = 0.0006) and non-peripheral (12.44 ± 5.94 vs 7.71 ± 7.1, p = 0.0183) groups. There was no significant difference in TBR of dbPET between the peripheral and non-peripheral groups.

CONCLUSION

The phantom study revealed poorer image quality at < 2-cm distance from the detector's edge than at other more central parts. In clinical studies, however, the visibility of breast lesions with dbPET was the same regardless of the lesion position, and it was higher than that in PET/CT. dbPET has a great potential for detecting breast lesions near the chest wall if they are at least 2 cm from the edge of the FOV, even in young women with small breasts.

摘要

背景

我们利用俯卧位悬垂乳房成像的体模和临床研究,评估了一款商用专用乳腺PET(dbPET)在探测器边缘(即胸壁附近乳腺所在位置)的图像质量,并将其与相同临床受试者的仰卧位PET/CT乳房图像进行比较。

方法

一个装有四个球体(直径分别为16毫米、10毫米、7.5毫米和5毫米)的乳房体模,填充氟脱氧葡萄糖溶液(球体与背景的活度浓度比为8:1)。球体从探测器的顶部边缘到中心占据五个不同位置,每个位置扫描5分钟。对重建图像进行视觉评估,并计算所有球体的对比噪声比(CNR)、对比恢复系数(CRC)以及背景的变异系数(CV)。随后,对标准仰卧位PET/CT和俯卧位dbPET获得的临床图像进行回顾性分析。比较胸壁附近(靠近探测器边缘;外周组)和其他位置(非外周组)乳腺癌之间的肿瘤与背景比值(TBR)。比较dbPET和PET/CT中每个病变的TBR。

结果

在体模研究中,离探测器边缘越近,所有球体的CNR和CRC降低,而CV增加。这种放置方式的缺点在视觉上得到了证实。关于临床图像,dbPET在外周组(12.38±6.41对6.73±3.5,p = 0.0006)和非外周组(12.44±5.94对7.71±7.1,p = 0.0183)的TBR均显著高于PET/CT。外周组和非外周组dbPET的TBR无显著差异。

结论

体模研究显示,距离探测器边缘小于2厘米处的图像质量比其他更中心部位差。然而,在临床研究中,无论病变位置如何,dbPET对乳房病变的可视性相同,且高于PET/CT。即使对于乳房较小的年轻女性,若胸壁附近的乳房病变距离视野边缘至少2厘米,dbPET在检测这些病变方面具有很大潜力。