视野外放射性对专用乳腺正电子发射断层成像质量的影响:初步的体模和临床研究。

Effect of radioactivity outside the field of view on image quality of dedicated breast positron emission tomography: preliminary phantom and clinical studies.

机构信息

Yamanashi PET Imaging Clinic, Chuo City, Yamanashi Prefecture, Japan.

Department of Radiology, University of Yamanashi, Chuo City, Yamanashi Prefecture, 409-3821, Japan.

出版信息

Ann Nucl Med. 2022 Dec;36(12):1010-1018. doi: 10.1007/s12149-022-01789-7. Epub 2022 Oct 8.

Abstract

OBJECTIVE

Semi-quantitative positron emission tomography (PET) values, such as the maximum standardized uptake value (SUV), are widely used to identify malignant lesions and evaluate the response to treatment. The image quality of ring-shaped dedicated breast positron emission tomography (dbPET) has been known to decrease the closer it is to the detector's edge. This study aimed to investigate the effect of radioactivity (RI) outside the detector field of view (FOV) on the image quality of the ring-shaped dbPET.

METHODS

A breast phantom containing the left myocardium, which was prepared using a 3D printer, filled with F-fluorodeoxyglucose (FDG) solution with various RI concentration ratios (RCRs) of myocardium to background and scanned with the edge of an apex positioned exactly in line with the edge of the FOV of the dbPET scanner. The phantom image quality was visually and quantitatively evaluated. Following the phantom study, left-right breast differences (the left breast uptake ratio to the right breast (LUR)) on clinical dbPET images of 74 women were quantitatively evaluated. The relationships between these parameters, clinical indices, and FDG uptake in the left myocardium on PET/computed tomography (CT) images were analyzed.

RESULTS

The phantom study showed that the higher the RCR of the myocardium and the closer it is to the top edge of the phantom, the higher is the pixel value of the dbPET images. In a clinical study, LUR was significantly correlated with myocardial SUV (r = 0.96, p < 0.0001) and metabolic myocardial volume (r = 0.63, p = 0.001) for whole-body PET/CT imaging. Although no significant correlations were found between LUR and age (r = 0.05, p = 0.6865), body mass index (r = 0.03, p = 0.8178), or distance between the left myocardial apex and chest wall (r = 0.16, p = 0.1667).

CONCLUSIONS

FDG uptake in the myocardium affected dbPET images of the left breast, especially near the chest wall. Further, the effect of RI outside the FOV, such as in the myocardium, must be considered in the quantitative evaluation of breast cancer using dbPET.

摘要

目的

半定量正电子发射断层扫描(PET)值,如最大标准化摄取值(SUV),被广泛用于识别恶性病变并评估治疗反应。已经知道,环形专用乳腺正电子发射断层扫描(dbPET)的图像质量随着其靠近探测器边缘而降低。本研究旨在探讨探测器视野(FOV)外的放射性(RI)对环形 dbPET 图像质量的影响。

方法

使用 3D 打印机制备包含左心室的乳腺体模,该体模用 F-氟脱氧葡萄糖(FDG)溶液填充,其左心室与背景的 RI 浓度比(RCR)具有不同的比例,并使用正好与 dbPET 扫描仪 FOV 边缘对齐的顶点位置对体模进行扫描。对体模图像进行了视觉和定量评估。在体模研究之后,对 74 名女性的临床 dbPET 图像进行了左-右乳腺差异(左乳腺摄取比右侧乳腺(LUR))的定量评估。分析了这些参数与临床指标以及 PET/CT(CT)图像上左心室 FDG 摄取之间的关系。

结果

体模研究表明,左心室的 RCR 越高,距离体模顶部越近,dbPET 图像的像素值越高。在临床研究中,LUR 与全身 PET/CT 成像的左心室 SUV(r=0.96,p<0.0001)和代谢性心肌体积(r=0.63,p=0.001)显著相关。尽管 LUR 与年龄(r=0.05,p=0.6865)、体重指数(r=0.03,p=0.8178)或左心室顶点与胸壁之间的距离(r=0.16,p=0.1667)之间没有显著相关性。

结论

左心室 FDG 摄取会影响左乳腺的 dbPET 图像,尤其是靠近胸壁的区域。此外,在使用 dbPET 对乳腺癌进行定量评估时,必须考虑到 FOV 外的 RI,如左心室的 RI。