IRE 治疗中导航引导下电极放置的准确性。

Accuracy of Electrode Placement in IRE Treatment with Navigated Guidance.

机构信息

Division of Surgery, Department of Clinical Sciences, Karolinska Institutet at Danderyd Hospital, Stockholm, Sweden.

Department of Surgery and Urology, Danderyd Hospital, 182 88, Stockholm, Sweden.

出版信息

Cardiovasc Intervent Radiol. 2021 Jun;44(6):968-975. doi: 10.1007/s00270-020-02762-5. Epub 2021 Jan 20.

Abstract

PURPOSE

Evaluate the accuracy of multiple electrode placements in IRE treatment of liver tumours using a stereotactic CT-based navigation system.

METHOD

Analysing data from all IRE treatments of liver tumours at one institution until 31 December 2018. Comparing planned with validated electrode placement. Analysing lateral and angular errors and parallelism between electrode pairs RESULTS: Eighty-four tumours were treated in 60 patients. Forty-six per cent were hepatocellular carcinoma, and 36% were colorectal liver metastases. The tumours were located in all segments of the liver. Data were complete from 51 treatments. Two hundred and six electrodes and 336 electrode pairs were analysed. The median lateral and angular error, comparing planned and validated electrode placement, was 3.6 mm (range 0.2-13.6 mm) and 3.1° (range 0°-16.1°). All electrodes with a lateral error >10 mm were either re-positioned or excluded before treatment. The median angle between the electrode pairs was 3.8° (range 0.3°-17.2°). There were no electrode placement-related complications.

CONCLUSION

The use of a stereotactic CT-based system for navigation of electrode placement in IRE treatment of liver tumours is safe, accurate and user friendly.

摘要

目的

评估使用基于立体定向 CT 的导航系统对肝脏肿瘤进行IRE 治疗时,多个电极放置的准确性。

方法

分析截至 2018 年 12 月 31 日,在一个机构中进行的所有肝脏肿瘤 IRE 治疗的数据。比较计划与验证的电极放置。分析电极对之间的侧向和角度误差以及平行度。

结果

在 60 名患者中治疗了 84 个肿瘤。46%为肝细胞癌,36%为结直肠肝转移。肿瘤位于肝脏的所有节段。51 次治疗中有完整的数据。分析了 206 个电极和 336 对电极。计划与验证的电极放置相比,中位数的侧向和角度误差分别为 3.6 毫米(范围 0.2-13.6 毫米)和 3.1°(范围 0°-16.1°)。所有侧向误差大于 10 毫米的电极均在治疗前重新定位或排除。电极对之间的中位数角度为 3.8°(范围 0.3°-17.2°)。没有与电极放置相关的并发症。

结论

使用基于立体定向 CT 的系统对肝脏肿瘤 IRE 治疗中的电极放置进行导航是安全、准确且易于使用的。