Objective Assessment of the Quality and Accuracy of Deformable Image Registration.

作者信息

Jurkovic Ines-Ana, Papanikolaou Nikos, Stathakis Sotirios, Kirby Neil, Mavroidis Panayiotis

机构信息

Department of Radiation Oncology, University of Texas Health Science Center at San Antonio, San Antonio, TX, USA.

Department of Radiation Oncology, University of North Carolina, Chapel Hill, North Carolina, USA.

出版信息

J Med Phys. 2020 Jul-Sep;45(3):156-167. doi: 10.4103/jmp.JMP_47_19. Epub 2020 Oct 13.

Abstract

BACKGROUND

The increased use of deformable registration algorithms in clinical practice has also increased the need for their validation.

AIMS AND OBJECTIVES

The purpose of the study was to investigate the quality, accuracy, and plausibility of three commercial image registration algorithms for 4-dimensional computed tomography (4DCT) datasets using various similarity measures.

MATERIALS AND METHODS

4DCT datasets were acquired for 10 lung cancer patients. 23 similarity measures were used to evaluate image registration quality. To ensure selected method's invertibility and assess resultant mechanical stress, the determinant of the Jacobian for the displacement field and 3-D Eulerian strain tensor were calculated. All the measures and calculations were applied on to extended deformable multi pass (EXDMP) and deformable multi pass (DMP) methods.

RESULTS

The results indicate the same trend for several of the studied measures. The Jacobian determinant values were always positive for the DMP method. The Eulerian strain tensor had smaller values for the DMP method than EXDMP in all of the studied cases. The negative values of the Jacobian determinant point to non-physical behavior of the EXDMP method. The Eulerian strain tensor values indicate less tissue strain for the DMP method. Large differences were also observed in the results between complete and cropped datasets (coefficient of determination: 0.55 vs. 0.93).

CONCLUSION

A number of error and distance measures showed the best performance among the tested measures. The evaluated measures might detect CT dataset differences with higher precision if the analysis is restricted to a smaller volume.

摘要

背景

临床实践中可变形配准算法使用的增加也加大了对其进行验证的需求。

目的

本研究旨在使用各种相似性度量方法,调查三种用于四维计算机断层扫描(4DCT)数据集的商业图像配准算法的质量、准确性和合理性。

材料与方法

获取了10名肺癌患者的4DCT数据集。使用23种相似性度量方法评估图像配准质量。为确保所选方法的可逆性并评估产生的机械应力,计算了位移场雅可比行列式和三维欧拉应变张量。所有度量和计算均应用于扩展可变形多步(EXDMP)和可变形多步(DMP)方法。

结果

对于几种研究的度量方法,结果显示出相同趋势。DMP方法的雅可比行列式值始终为正。在所有研究案例中,DMP方法的欧拉应变张量值均小于EXDMP方法。雅可比行列式的负值表明EXDMP方法存在非物理行为。欧拉应变张量值表明DMP方法的组织应变较小。完整数据集和裁剪数据集的结果之间也观察到较大差异(决定系数:0.55对0.93)。

结论

一些误差和距离度量方法在测试的度量方法中表现最佳。如果将分析限制在较小体积内,所评估的度量方法可能会以更高精度检测CT数据集差异。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/38f7/7810144/11bac3df5a51/JMP-45-156-g001.jpg