Suppr超能文献

增强现实引导下软组织穿刺活检:一项初步研究。

Augmented Reality Guided Needle Biopsy of Soft Tissue: A Pilot Study.

作者信息

Asgar-Deen David, Carriere Jay, Wiebe Ericka, Peiris Lashan, Duha Aalo, Tavakoli Mahdi

机构信息

Electrical and Computer Engineering, University of Alberta, Edmonton, AB, Canada.

Oncology, Medicine and Dentistry, University of Alberta, Edmonton, AB, Canada.

出版信息

Front Robot AI. 2020 Jun 16;7:72. doi: 10.3389/frobt.2020.00072. eCollection 2020.

Abstract

Percutaneous biopsies are popular for extracting suspicious tissue formations (primarily for cancer diagnosis purposes) due to the: relatively low cost, minimal invasiveness, quick procedure times, and low risk for the patient. Despite the advantages provided by percutaneous biopsies, poor needle and tumor visualization is a problem that can result in the clinicians classifying the tumor as benign when it was malignant (false negative). The system developed by the authors aims to address the concern of poor needle and tumor visualization through two virtualization setups. This system is designed to track and visualize the needle and tumor in three-dimensional space using an electromagnetic tracking system. User trials were conducted in which the 10 participants, who were not medically trained, performed a total of 6 tests, each guiding the biopsy needle to the desired location. The users guided the biopsy needle to the desired point on an artificial spherical tumor (diameters of 30, 20, and 10 mm) using the 3D augmented reality (AR) overlay for three trials and a projection on a second monitor (TV) for the other three trials. From the randomized trials, it was found that the participants were able to guide the needle tip 6.5 ± 3.3 mm away from the desired position with an angle deviation of 1.96 ± 1.10° in the AR trials, compared to values of 4.5 ± 2.3 mm and 2.70 ± 1.67° in the TV trials. The results indicate that for simple stationary surgical procedures, an AR display is non-inferior a TV display.

摘要

经皮活检因成本相对较低、侵入性极小、操作时间短以及患者风险低等优点,在提取可疑组织形成物(主要用于癌症诊断)方面很受欢迎。尽管经皮活检有诸多优势,但针和肿瘤的可视化不佳是个问题,可能导致临床医生将恶性肿瘤误诊为良性(假阴性)。作者开发的系统旨在通过两种虚拟设置解决针和肿瘤可视化不佳的问题。该系统设计使用电磁跟踪系统在三维空间中跟踪和可视化针与肿瘤。进行了用户试验,10名未接受医学培训的参与者总共进行了6次测试,每次都将活检针引导到所需位置。在三次试验中,用户使用3D增强现实(AR)叠加将活检针引导到人工球形肿瘤(直径分别为30、20和10毫米)上的所需点,另外三次试验则引导到第二个显示器(电视)上的投影处。从随机试验中发现,在AR试验中,参与者能够将针尖引导至距离所需位置6.5±3.3毫米处,角度偏差为1.96±1.10°,而在电视试验中的值分别为4.5±2.3毫米和2.70±1.67°。结果表明,对于简单的固定外科手术,AR显示器并不逊于电视显示器。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。