Byvaltsev Vadim A, Kalinin Andrei A, Aliyev Marat A, Riew K Daniel
64877Irkutsk State Medical University, Irkutsk, Russia.
Railway Clinical Hospital, Irkutsk, Russia.
Global Spine J. 2023 Jan;13(1):67-73. doi: 10.1177/2192568221991105. Epub 2021 Jan 28.
Prospective non-randomized single-center cohort study.
To analyze the quality of postoperative magnetic resonance imaging of 2 structurally different cervical disc arthroplasty devices at the index and adjacent levels.
A non-randomized, comparative, prospective, single-center study included 40 patients (23 men and 17 women) aged 32 (26-40) years. Two study groups were utilized: in the first (n = 20), a titanium prosthesis was used; in the second (n = 20), a cobalt-chromium implant was used. Evaluation of MRI studies before and after surgery was performed using sagittal and axial T2 weighted images by 2 specialists who were blinded to the prosthesis that was used. To determine the quality of an MRI image, the classification of Jarvik 2000, the radiological and orthopedic scales for assessing artifacts were used.
There was good-to-excellent inter-observer agreement for all of the MR parameters used for the titanium and satisfactory-to-good for the cobalt chromium group. The analysis of the quality of postoperative imaging using the Jarvik 2000 scale showed a statistically significant deterioration in MR images in the cobalt chromium group ( < 0.001), compared to the titanium ( = 0.091). Following a single-level total arthroplasty, the titanium group had better MRI images according to radiological and orthopedic scales ( < 0.001).
Titanium cervical disc arthroplasty devices result in superior postoperative MR imaging, as compared to cobalt chromium prostheses, as the latter significantly reduces image quality due to the pronounced ferromagnetic effect.
前瞻性非随机单中心队列研究。
分析两种结构不同的颈椎间盘置换装置在手术节段及相邻节段术后磁共振成像的质量。
一项非随机、对比、前瞻性、单中心研究纳入了40例年龄在32(26 - 40)岁的患者(23例男性和17例女性)。使用了两个研究组:第一组(n = 20)使用钛制假体;第二组(n = 20)使用钴铬合金植入物。由两名对所使用的假体不知情的专家,通过矢状面和轴位T2加权图像对手术前后的磁共振成像研究进行评估。为确定磁共振图像的质量,使用了Jarvik 2000分类法以及用于评估伪影的放射学和矫形学量表。
对于钛制假体所使用的所有磁共振参数,观察者间一致性良好至优秀;对于钴铬合金组,观察者间一致性满意至良好。使用Jarvik 2000量表对术后成像质量进行分析显示,与钛制假体组(P = 0.091)相比,钴铬合金组的磁共振图像有统计学意义的恶化(P < 0.001)。在进行单节段全关节置换术后,根据放射学和矫形学量表,钛制假体组的磁共振图像更好(P < 0.001)。
与钴铬合金假体相比,钛制颈椎间盘置换装置术后磁共振成像效果更佳,因为后者由于明显的铁磁效应会显著降低图像质量。