Department of Medical Biochemistry, Faculty of Medicine, Gaziantep University, Gaziantep, Turkey.
Int J Lab Hematol. 2021 Aug;43(4):616-622. doi: 10.1111/ijlh.13444. Epub 2021 Feb 1.
With recent advances in technology, many manual tests are being replaced by automated devices due to a wide range of advantages. One of these tests is the erythrocyte sedimentation rate (ESR) test that is used to determine inflammatory activity. This study aimed to evaluate the agreement between the Starrsed Interliner sedimentation device and the gold standard method, that is the Westergren method, used in ESR measurement.
One hundred fifty-one patients who presented to Gaziantep University Faculty of Medicine, Şahinbey Training and Research Hospital were included in this study. ESR values were measured simultaneously within 2 hours using the ESR analyzer Starrsed Interliner device and the gold standard method of measuring ESR, that is the Westergren method, from blood samples collected from the same patients in EDTA and citrate tubes.
Agreement between the results from the Starrsed Interliner device and the Westergren method was evaluated using the Intraclass Correlation method. Consequently, a poor correlation was observed at values <20 mm/h, a moderate correlation was observed at values 20 to 80 mm/h and >80 mm/h, and an excellent correlation was observed when all results were considered. Method comparison was conducted according to the Passing-Bablok regression analysis (y = -1.50 + 0.75x) (P < .0001). The mean difference between the two methods was 10.1 according to the Bland-Altman analysis.
Despite the advantages of the Starrsed Interliner device, such as lower laboratory workloads, lower costs and turnaround time, the difference between the two methods, as found in this study, may lead to different clinical interpretations for results in some patient.
随着技术的最新进展,由于具有广泛的优势,许多手动测试正被自动化设备所取代。这些测试之一是红细胞沉降率(ESR)测试,用于确定炎症活动。本研究旨在评估 Starrsed Interliner 沉降设备与 ESR 测量的金标准方法(即 Westergren 方法)之间的一致性。
本研究纳入了 151 名就诊于加济安泰普大学医学院、沙欣贝伊培训与研究医院的患者。使用 EDTA 和柠檬酸盐管从同一患者采集的血液样本,在 2 小时内,使用 ESR 分析仪 Starrsed Interliner 设备和 ESR 测量的金标准方法(即 Westergren 方法)同时测量 ESR 值。
使用组内相关法评估 Starrsed Interliner 设备与 Westergren 方法的结果之间的一致性。结果表明,在<20mm/h 时观察到相关性较差,在 20 至 80mm/h 和>80mm/h 时观察到中度相关性,当考虑所有结果时观察到极好的相关性。根据 Passing-Bablok 回归分析(y=-1.50+0.75x)(P<.0001)进行方法比较。根据 Bland-Altman 分析,两种方法之间的平均差异为 10.1。
尽管 Starrsed Interliner 设备具有降低实验室工作量、降低成本和周转时间等优势,但本研究发现两种方法之间的差异可能导致某些患者的结果产生不同的临床解释。