Wang Yanyun, Sun Yun, Jiang Tao
Nanjing Maternity and Child Health Care Hospital, Women's Hospital of Nanjing Medical University, Nanjing, China.
Front Pediatr. 2021 Jan 21;8:596321. doi: 10.3389/fped.2020.596321. eCollection 2020.
Tandem mass spectrometry becomes a common and important test in newborn screening, but potential contamination of the equipment has largely been ignored. The source of contamination through Biosan quality control samples was examined prospectively, and further confirmed by retrospective analysis of patient samples. We found that the source of contamination came from a syringe in the Biosan quality control samples. Furthermore, we found that a large number of indicators in the patient sample were interfered by syringe contamination in our center, and also in two other newborn screening centers, but the affected indicators were different in different screening centers. Syringe contamination will affect the detection of patient samples by tandem mass spectrometry and should be monitored carefully and immediately.
串联质谱法已成为新生儿筛查中一项常见且重要的检测方法,但设备的潜在污染在很大程度上被忽视了。前瞻性地检查了通过Biosan质量控制样本造成污染的来源,并通过对患者样本的回顾性分析进一步证实。我们发现污染来源来自Biosan质量控制样本中的一支注射器。此外,我们发现,在我们中心以及另外两个新生儿筛查中心,患者样本中的大量指标受到注射器污染的干扰,但不同筛查中心受影响的指标不同。注射器污染会影响串联质谱法对患者样本的检测,应立即进行仔细监测。