Abou-Ayash Samir, Schimmel Martin, Worni Andreas, Enkling Norbert
Int J Prosthodont. 2023 Jan-Feb;36(1):59–62. doi: 10.11607/ijp.7055.
To analyze the influence of pristine matrix and O-ring dimensions on retention force and reproducibility in single one-piece mini dental implants (MDIs) with ball patrices under in vitro conditions.
Three different matrix and O-ring combinations (MH1-MH3) were evaluated (n = 50 per group) on 1.8-mm-diameter implants. The matrices were manually mounted on the implants and were subsequently removed in a vertical linear manner using a metal pin with two strain gauges, recording the maximum force during disconnection. After five disconnections, the O-rings were exchanged, and the mean retention force was calculated, resulting in 50 values for each matrix and O-ring combination. Mean retention forces, SDs, and 95% CI were calculated. Analysis of variance was used to test the global differences, and post hoc pairwise comparisons were subsequently applied. The level of significance was set to P < .05.
ANOVA (global P < .0001) and pairwise comparisons (all P < .0001) demonstrated statistically significant differences among the three different matrix and O-ring combinations, with mean values of 5.18 N (MH 1), 6.73 (MH 2), and 9.08 (MH 3). Within each combination, retention force variations of > 1 N could not be demonstrated; ie, by exchanging O-rings, a similar retention force can be reestablished.
Matrix and O-ring dimensions have a significant influence on retention forces in one-piece MDIs. Pristine O-rings demonstrated highly reproducible initial retention forces in all matrices.
分析在体外条件下,原始基质和O形环尺寸对带有球形基体的单件式微型牙种植体(MDIs)固位力和可重复性的影响。
在直径为1.8毫米的种植体上评估了三种不同的基质和O形环组合(MH1 - MH3)(每组n = 50)。将基质手动安装在种植体上,随后使用带有两个应变片的金属销以垂直线性方式将其移除,记录断开过程中的最大力。在五次断开后,更换O形环,并计算平均固位力,每种基质和O形环组合得到50个值。计算平均固位力、标准差和95%置信区间。采用方差分析来检验总体差异,随后进行事后两两比较。显著性水平设定为P < 0.05。
方差分析(总体P < 0.0001)和两两比较(所有P < 0.0001)表明,三种不同的基质和O形环组合之间存在统计学上的显著差异,平均值分别为5.18 N(MH1)、6.73 N(MH2)和9.08 N(MH3)。在每种组合内,未发现固位力变化> 1 N的情况;即,通过更换O形环,可以重新建立相似的固位力。
基质和O形环尺寸对单件式MDIs的固位力有显著影响。原始O形环在所有基质中均显示出高度可重复的初始固位力。