J Vet Diagn Invest. 2021 May;33(3):611-614. doi: 10.1177/1040638721998295. Epub 2021 Mar 1.
Refractometry is utilized routinely to evaluate canine urine specific gravity (USG) in veterinary clinical settings. We aimed to determine if the magnitude of interobserver reliability when assessing canine USG via refractometry could impact clinical judgment. USG was determined in 38 dogs by 3 registered veterinary technicians (RVTs) using both an optical analog refractometer and a digital refractometer. Summary statistics were reported, interobserver reliability was assessed via intraclass correlation coefficient (ICC) analysis through a 2-way mixed-effects model, and agreement between RVT pairs was compared through Bland-Altman plots. The median analog refractometer USG measurement was 1.018 (range: 1.004-1.040) and for the digital refractometer was 1.0176 (1.0035-1.0357). The analog refractometer average measure ICC was 0.995 (95% CI: 0.992, 0.997; < 0.001). The digital refractometer average measure ICC was 0.999 (95% CI: 0.999, 1.000; < 0.001). Strong agreement between all pairs of RVTs was seen via Bland-Altman plots for both analog and digital refractometers, with 95% CIs spanning no more than 0.002 in either the positive or negative direction for all pairings. The interobserver variability in canine USG measurements by RVTs was trivial and did not impact clinical judgment and decision-making.
折射仪常用于兽医临床环境中评估犬尿液比重(USG)。我们旨在确定评估犬 USG 时的观察者间可靠性的幅度是否会影响临床判断。通过 3 名注册兽医技术员(RVT)使用光学模拟折射仪和数字折射仪,对 38 只狗的 USG 进行了测定。报告了统计摘要,通过双向混合效应模型的组内相关系数(ICC)分析评估了观察者间的可靠性,并通过 Bland-Altman 图比较了 RVT 对之间的一致性。模拟折射仪 USG 测量的中位数为 1.018(范围:1.004-1.040),数字折射仪为 1.0176(1.0035-1.0357)。模拟折射仪的平均测量 ICC 为 0.995(95%CI:0.992,0.997; < 0.001)。数字折射仪的平均测量 ICC 为 0.999(95%CI:0.999,1.000; < 0.001)。通过 Bland-Altman 图,所有 RVT 对之间都显示出强烈的一致性,在正向或负向方向上,所有配对的 95%CI 均不超过 0.002。RVT 对犬 USG 测量的观察者间变异性微不足道,不会影响临床判断和决策。