Mount J N, Kearney E M, Rosseneu M, Slavin B M
Chemical Pathology and Metabolic Disorders, St Thomas's Hospital, London.
J Clin Pathol. 1988 Apr;41(4):471-4. doi: 10.1136/jcp.41.4.471.
Immunoturbidimetric assays for measuring the apolipoproteins A1 and B using the Cobas Bio centrifugal analyser are described. The methods were specific, offered good sensitivity (less than 0.05 g/l) and intrabatch variability, with coefficients of variation between 2.4% and 3.5%, and were cost effective. Reference ranges were calculated for a group of civil servants, aged 35 to 55 years.
本文描述了使用Cobas Bio离心分析仪通过免疫比浊法测定载脂蛋白A1和B的方法。这些方法具有特异性,灵敏度高(低于0.05 g/l),批内变异小,变异系数在2.4%至3.5%之间,且成本效益高。还计算了一组年龄在35至55岁之间的公务员的参考范围。