Greene E R, Caprihan A, Davis J G
Lovelace Medical Foundation.
ISA Trans. 1988;27(1):43-50.
We developed sample volume tracking units that controlled the pulsed ultrasonic Doppler sample volume location in relation with the anterior-posterior movement of the human left main coronary artery (LMCA). Combined with noninvasive Doppler echocardiographic mechanical sector scanners (DS), the trackers controlled the axial location of the sample volume by range gate control. The Doppler angle was minimized with the long axis of the imaged LMCA. Both stored waveform, memory-driven (MD) and real-time (RT) trackers were developed. These devices were used to measure blood velocity spectral waveforms and lumen diameters, which were used to calculate flow. Using the RT tracker, we compared DS measurements with known flows (0-1000 ml/min) in a moving 4 mm tygon tubing phantom (r = 0.92, SEE = 32 ml/min). Using the MD tracker, we compared the DS with simultaneous invasive flow measurements in 11 patients with angiographically normal coronaries and ventricular function during cardiac catheterization. Using the RT tracker, we compared the DS measurements with subsequent nonsimultaneous, invasive flow data in 8 similar patients. Invasively determined flows were calculated from angiographic diameters and blood velocities which were obtained with a Doppler velocimeter catheter. Regression coefficients (r) were: (Table: see text). V = maximum velocity, D = diameter, Q = mean flow, *p less than .05 Interoperator and intraoperator variabilities in vivo of DS measurements with RT tracking were 21% and 15%, respectively. We conclude that Doppler echocardiography with either MD or RT sample volume tracking may be of limited clinical usefulness in the noninvasive measurement of phasic left main coronary artery blood flow in unselected patients.
我们开发了样本容积跟踪装置,该装置可根据人体左冠状动脉(LMCA)前后移动来控制脉冲超声多普勒样本容积的位置。这些跟踪装置与无创多普勒超声心动图机械扇形扫描仪(DS)相结合,通过距离选通控制来控制样本容积的轴向位置。使多普勒角度与所成像的LMCA长轴最小化。我们开发了存储波形、内存驱动(MD)和实时(RT)两种跟踪装置。这些装置用于测量血流速度频谱波形和管腔直径,进而计算血流量。使用实时跟踪装置,我们在一个移动的4毫米泰科管模型中(r = 0.92,标准估计误差 = 32毫升/分钟),将DS测量结果与已知流量(0 - 1000毫升/分钟)进行比较。使用内存驱动跟踪装置,我们在11例冠状动脉造影正常且心室功能正常的患者进行心导管检查期间,将DS测量结果与同步有创血流测量结果进行比较。使用实时跟踪装置,我们在8例类似患者中将DS测量结果与随后的非同步有创血流数据进行比较。有创测定的流量由血管造影直径和使用多普勒测速导管获得的血流速度计算得出。回归系数(r)为:(表格:见正文)。V = 最大速度,D = 直径,Q = 平均流量,*p小于0.05。使用实时跟踪时,DS测量在体内的操作者间和操作者内变异性分别为21%和15%。我们得出结论,对于未经过筛选的患者,使用内存驱动或实时样本容积跟踪的多普勒超声心动图在无创测量左冠状动脉阶段性血流方面的临床实用性可能有限。