Dingeon B, Charvin M A, Quenard M T, Thome H
Service de Biologie, Centre Hospitalier Chambery, France.
Clin Chem. 1988 Jun;34(6):1119-21.
Measurement of acetaminophen by analysis of the second derivative of its spectrum is specific and sensitive. The method of extraction and the use of just one phosphate buffer as reagent makes this method very convenient. Readings are reliable from 10 to 1500 mg/L. A turnaround time of 20 min makes this method well suited for emergency cases. Precision and accuracy of the method are presented. Results are not biased by interferences, not even from N-acetylcysteine.
通过对扑热息痛光谱的二阶导数进行分析来测定扑热息痛,该方法具有特异性和灵敏性。提取方法以及仅使用一种磷酸盐缓冲液作为试剂使得该方法非常便捷。在10至1500毫克/升的范围内读数可靠。20分钟的周转时间使该方法非常适合紧急情况。文中给出了该方法的精密度和准确度。结果不受干扰因素的影响,即使是来自N-乙酰半胱氨酸的干扰也不会产生偏差。