豚鼠眼 A 扫描超声与 Lenstar 光学生物测量仪的比较。

Comparison of A-Scan ultrasonography and the Lenstar optical biometer in Guinea pig eyes.

机构信息

Tianjin Key Laboratory of Retinal Functions and Diseases, Tianjin International Joint Research and Development Centre of Ophthalmology and Vision Science, Eye Institute and School of Optometry, Tianjin Medical University Eye Hospital, Tianjin, China.

Department of Physiology, University of Oklahoma Health Sciences Center, Oklahoma, USA.

出版信息

Exp Eye Res. 2021 Jun;207:108578. doi: 10.1016/j.exer.2021.108578. Epub 2021 Apr 20.

Abstract

OBJECTIVE

To compare the biometric parameters provided by A-scan ultrasonography and the Lenstar optical biometer in guinea pig eyes, including anterior segment depth (ASD), lens thickness (LT), vitreous chamber depth (VCD), and axial length (AL), and differences of them between treated form deprivation (FD) eyes and untreated fellow eyes after 4 weeks of FD.

METHODS

Three-week-old guinea pigs (N = 41) were subjected to biometric measurements before monocular FD (baseline) and after a 4-week FD. Statistical analyses including within-subject standard deviation (SD), coefficient of variation (CV), and intraclass correlation coefficient (ICC), used to evaluate repeatability for both the A-scan ultrasonography and the Lenstar individually, and correlation and Bland-Altman analyses were used to assess agreement between the two methods. The absolute values of ASD, LT, VCD and AL as measured by the two devices were compared, and the differences of them between treated (T) and untreated fellow (F) eyes (ΔASD, ΔLT, ΔVCD and ΔAL) (Δ = T-F) were compared between the two devices after 4 weeks of FD.

RESULTS

Measurements by the Lenstar (ICC: 0.923-0.994) were more repeatable than A-scan ultrasonography (ICC: 0.825-0.870). There was a high correlation for AL (r = 0.851, P < 0.001), a moderate correlation for VCD (r = 0.571, P < 0.001) and LT (r = 0.423, P < 0.001), and a low correlation for ASD (r = 0.230, P < 0.01) between the two devices. The values for ASD, VCD and AL measured by A-scan ultrasonography were larger than those measured by the Lenstar (all, P < 0.001), while LT provided by A-scan ultrasonography was much smaller than that of the Lenstar (P < 0.001). Bland-Altman plots showed poor agreement of absolute values of the four parameters between the two devices. Moreover, there was a high correlation between both methods for ΔAL (r = 0.704, P < 0.001), a moderate correlation for ΔVCD (r = 0.534, P < 0.001) and ΔASD (r = 0.574, P < 0.001), and no correlation for ΔLT (r = 0.303, P = 0.054). The ΔASD, ΔLT, and ΔAL measurements obtained by A-scan ultrasonography were greater than those obtained by the Lenstar (all, P < 0.001), while ΔVCD was mildly smaller using A-scan ultrasonography (P < 0.05). Bland-Altman plots illustrated there is good agreement of ΔAL, ΔVCD, ΔASD, and ΔLT between the two devices.

CONCLUSIONS

The Lenstar exhibited better repeatability and provided smaller measurements for AL, VCD and ASD than A-scan ultrasonography. Furthermore, a high correlation and a good agreement for the ΔAL was observed between the two devices after a period of FD. In summary, the two devices cannot replace each other directly to obtain absolute values of ASD, LT, VCD and AL, but the Lenstar still can serve as an option in measuring ΔAL between eyes in guinea pig myopia model.

摘要

目的

比较 A 型超声和 Lenstar 光学生物测量仪在豚鼠眼的生物测量参数,包括前节深度(ASD)、晶状体厚度(LT)、玻璃体腔深度(VCD)和眼轴长度(AL),以及 4 周形觉剥夺(FD)后治疗眼和未治疗对侧眼之间的差异。

方法

3 周龄豚鼠(N=41)在单眼 FD 前(基线)和 4 周 FD 后进行生物测量。使用组内标准差(SD)、变异系数(CV)和组内相关系数(ICC)进行统计分析,分别评估 A 型超声和 Lenstar 的重复性,以及两种方法之间的相关性和 Bland-Altman 分析。比较两种仪器测量的 ASD、LT、VCD 和 AL 的绝对值,并比较 4 周 FD 后两种仪器测量的治疗眼(T)和未治疗对侧眼(F)(ΔASD、ΔLT、ΔVCD 和 ΔAL)(Δ=T-F)之间的差异。

结果

Lenstar(ICC:0.923-0.994)的测量重复性优于 A 型超声(ICC:0.825-0.870)。AL 之间具有高度相关性(r=0.851,P<0.001),VCD 之间具有中度相关性(r=0.571,P<0.001)和 LT(r=0.423,P<0.001),ASD 之间具有低度相关性(r=0.230,P<0.01)。两种仪器之间 ASD、VCD 和 AL 的测量值,A 型超声测量值均大于 Lenstar(均 P<0.001),而 A 型超声测量的 LT 值明显小于 Lenstar(P<0.001)。Bland-Altman 图显示两种仪器 4 个参数的绝对值之间一致性较差。此外,两种方法之间的ΔAL 高度相关(r=0.704,P<0.001),ΔVCD 中度相关(r=0.534,P<0.001)和ΔASD 中度相关(r=0.574,P<0.001),ΔLT 之间无相关性(r=0.303,P=0.054)。A 型超声测量的ΔASD、ΔLT 和 ΔAL 测量值均大于 Lenstar(均 P<0.001),而使用 A 型超声测量的ΔVCD 略小(P<0.05)。Bland-Altman 图表明两种仪器之间的ΔAL、ΔVCD、ΔASD 和 ΔLT 具有良好的一致性。

结论

Lenstar 显示出更好的重复性,并为 AL、VCD 和 ASD 提供比 A 型超声更小的测量值。此外,在一段时间的 FD 后,两种设备之间的ΔAL 具有高度相关性和良好的一致性。总之,两种设备不能直接相互替代以获得 ASD、LT、VCD 和 AL 的绝对值,但 Lenstar 仍然可以作为豚鼠近视模型中测量对侧眼ΔAL 的一种选择。