Ferguson Jack, Dunn Steven, Best Angus, Mirza Jeremy, Percival Benita, Mayhew Megan, Megram Oliver, Ashford Fiona, White Thomas, Moles-Garcia Emma, Crawford Liam, Plant Tim, Bosworth Andrew, Kidd Michael, Richter Alex, Deeks Jonathan, McNally Alan
Institute of Cancer and Genomic Science, College of Medical and Dental Science, University of Birmingham, Birmingham, United Kingdom.
Institute of Microbiology and Infection, College of Medical and Dental Science, University of Birmingham, Birmingham, United Kingdom.
PLoS Biol. 2021 Apr 29;19(4):e3001216. doi: 10.1371/journal.pbio.3001216. eCollection 2021 Apr.
Lateral flow devices (LFDs) are quickly being implemented for use in large-scale population surveillance programs for SARS-CoV-2 infection in the United Kingdom. These programs have been piloted in city-wide screening in the city of Liverpool and are now being rolled out to support care home visits and the return home of University students for the Christmas break. Here, we present data on the performance of LFDs to test almost 8,000 students at the University of Birmingham between December 2 and December 9, 2020. The performance is validated against almost 800 samples using PCR performed in the University Pillar 2 testing lab and theoretically validated on thousands of Pillar 2 PCR testing results performed on low-prevalence care home testing samples. Our data show that LFDs do not detect infections presenting with PCR Ct values over 29 to 30 as determined using the Thermo Fisher TaqPath asssay. This may be of particular importance in detecting individuals that are either at the early, or late stages of infection, and reinforces the need for frequent, recurrent testing.
侧向流动检测设备(LFDs)正在迅速应用于英国针对新冠病毒感染的大规模人群监测项目。这些项目已在利物浦市进行全市范围的筛查试点,目前正在推广,以支持对养老院的探访以及大学生圣诞节假期返乡。在此,我们展示了2020年12月2日至12月9日期间在伯明翰大学对近8000名学生使用LFDs进行检测的性能数据。该性能通过在大学二级检测实验室进行的聚合酶链式反应(PCR)针对近800个样本进行了验证,并在对低流行率养老院检测样本进行的数千次二级PCR检测结果上进行了理论验证。我们的数据表明,使用赛默飞世尔TaqPath检测法时,LFDs无法检测出PCR Ct值超过29至30的感染情况。这在检测处于感染早期或晚期的个体时可能尤为重要,并强化了频繁、反复检测的必要性。