Colditz P B, Williams G L, Berry A B, Symonds P J
Department of Neonatology, Children's Hospital, Sydney, Australia.
Crit Care Med. 1988 Sep;16(9):876-9. doi: 10.1097/00003246-198809000-00011.
Fontanelle pressure (FP) and cerebral perfusion pressure (CPP) were measured continuously for the first 5 postnatal days in 21 preterm neonates. A new application device for use with the sensor of an intracranial pressure monitor allows accurate, continuous FP measurement. Accuracy of the technique was established by comparing results using the new device with 14-h continuous intraventricular pressure recordings from four infants before ventricular shunting. Agreement between the two techniques was satisfactory with the mean difference being 0.4 mm Hg. FP rose on each of the first 5 days after birth. CPP also rose over the 5 days but there were periods during suctioning of the endotracheal tube, crying, and vigorous activity when CPP was very low.
对21例早产新生儿出生后的前5天连续测量囟门压力(FP)和脑灌注压(CPP)。一种与颅内压监测器传感器配套使用的新型应用装置可实现准确、连续的FP测量。通过将使用新装置的测量结果与4例婴儿在脑室分流术前14小时的连续脑室内压力记录进行比较,确定了该技术的准确性。两种技术之间的一致性良好,平均差异为0.4毫米汞柱。出生后的前5天,FP每天都在升高。CPP在这5天中也有所升高,但在气管内插管抽吸、哭闹和剧烈活动期间,CPP会非常低。