CEA, CNRS, Inserm, BioMaps, Université Paris-Saclay, Orsay, France.
Laboratory of Imaging Biomarkers, Center for Research on Inflammation, UMR 1149, Inserm, Université de Paris, Paris, France.
NMR Biomed. 2021 Aug;34(8):e4543. doi: 10.1002/nbm.4543. Epub 2021 May 26.
In the framework of algebraic inversion, magnetic resonance elastography (MRE) repeatability, reproducibility and robustness were evaluated on extracted shear velocities (or elastic moduli). The same excitation system was implemented at two sites equipped with clinical MR scanners of 1.5 and 3 T. A set of four elastic, isotropic, homogeneous calibrated phantoms of distinct elasticity representing the spectrum of liver fibrosis severity was mechanically characterized. The repeatability of the measurements and the reproducibility between the two platforms were found to be excellent with mean coefficients of variations of 1.62% for the shear velocity mean values and 1.95% for the associated standard deviations. MRE velocities were robust to the amplitude and pattern variations of the displacement field with virtually no difference between outcomes from both magnets at identical excitation frequencies, even when the displacement field amplitude was six times smaller. However, MRE outcomes were very sensitive to the number of voxels per wavelength, s, of the recorded displacement field, with relative biases reaching 62% and precision loss by a factor of up to 23.5. For both magnetic field strengths, MRE accuracy and precision were largely degraded outside of established conditions of validity (6 ≲ s ≲ 9), resulting in estimated shear velocity values not significantly different between phantoms of increasing elasticity. When fulfilling the spatial sampling conditions, either prospectively in the acquisition or retrospectively before the reconstruction, MRE produced quantitative measurements that allowed to unambiguously discriminate, with infinitesimal p values, between the phantoms mimicking increasing severity of liver fibrosis.
在代数反演框架内,通过提取剪切速度(或弹性模量)评估了磁共振弹性成像(MRE)的重复性、可再现性和稳健性。在配备有 1.5T 和 3T 临床磁共振扫描仪的两个站点上实施了相同的激励系统。通过机械特性表征了一组具有不同弹性的四个弹性各向同性均匀校准的幻影,代表了肝纤维化严重程度的范围。发现测量的重复性和两个平台之间的可再现性非常好,剪切速度平均值的平均变异系数为 1.62%,相关标准差的平均变异系数为 1.95%。MRE 速度对位移场的幅度和模式变化具有稳健性,即使在相同的激励频率下,两个磁铁的结果几乎没有差异,即使位移场的幅度小了六倍。然而,MRE 结果对记录的位移场的每个波长的体素数 s 非常敏感,相对偏差达到 62%,精度损失高达 23.5 倍。对于两种磁场强度,MRE 的准确性和精密度在超出有效条件(6≲s≲9)时大大降低,导致在弹性不断增加的幻影之间,估计的剪切速度值没有显著差异。当满足空间采样条件时,无论是在采集过程中前瞻性地还是在重建前回顾性地,MRE 都能产生定量测量结果,能够以极小的 p 值明确区分模拟肝纤维化严重程度不断增加的幻影。