Adriano L Z, Barasuol J C, Cardoso M, Bolan M
Federal University of Santa Catarina (UFSC), Trindade, Florianópolis, SC, 88040-900, Brazil.
Federal University of Santa Catarina, João Pio Duarte Silva, 114-505 B, Córrego Grande, 88037-000, Florianópolis, SC, Brazil.
Eur Arch Paediatr Dent. 2019 Oct;20(5):403-408. doi: 10.1007/s40368-018-00413-5. Epub 2019 Jan 2.
To compare the accuracy of three electronic apex locators in the measurement of the root length of primary teeth with the radiographic and direct visual techniques.
The sample of the in vitro study was 102 root canals of primary molars. For the direct visual technique, the root canal file was introduced into the canal until its tip passed the apical foramen, then it was moved back to the limit of the foramen or root resorption and the measurement was made with a millimetre ruler. For the radiographic measurement, a calibrated and blinded researcher (κ > 0.87) measured the root canals using the same rule. Measurements with apex locators were performed by inserting the file into the canal until the word "Apex" or audible warning to indicate the foramen or resorption. For data analysis, one-way ANOVA for repeated measurements and the Sidak post hoc tests were performed.
The direct measurement (µ = 8.57) did not present statistical difference when compared to the Root ZX II measurement (µ = 8.45), as opposed to the radiographic measurements (µ = 8.12), the Endus (µ = 8.24) and iPex II (µ = 8.29).
The use of apex locators was superior to the radiographic method and the Root ZX II was the most effective.
比较三种电子根尖定位仪与影像学和直接视觉技术在测量乳牙牙根长度方面的准确性。
体外研究样本为102颗乳磨牙的根管。对于直接视觉技术,将根管锉插入根管,直至其尖端通过根尖孔,然后将其退回至根尖孔或牙根吸收的界限处,并用毫米尺进行测量。对于影像学测量,由一位经过校准且不知情的研究人员(κ>0.87)使用相同规则测量根管。使用根尖定位仪进行测量时,将锉插入根管,直至出现“Apex”字样或听到指示根尖孔或吸收的声音警告。进行数据分析时,采用重复测量的单因素方差分析和Sidak事后检验。
与Root ZX II测量值(μ=8.45)相比,直接测量值(μ=8.57)无统计学差异,而与影像学测量值(μ=8.12)、Endus(μ=8.24)和iPex II(μ=8.29)相比有统计学差异。
根尖定位仪的使用优于影像学方法,且Root ZX II最为有效。