Jean Mathieu, Schulmeister Karl
Seibersdorf Labor GmbH, Seibersdorf 2444, Austria.
Biomed Opt Express. 2021 Apr 5;12(5):2586-2603. doi: 10.1364/BOE.422618. eCollection 2021 May 1.
The exposure and emission limits of ICNIRP, IEC 60825-1 and ANSI Z136.1 to protect the skin are based on a limited number of studies. To broaden the database, a computer model was developed to predict injury thresholds in the wavelength range from 400 nm to 20 µm and was validated by comparison with all applicable experimental threshold data (ED) in the wavelength range from 488 nm to 10.6 µm and exposure durations between 8 µs and 630 s. The model predictions compare favorably with the data with an average ratio of computer prediction to ED of 1.01 (standard deviation ± 46%) and a maximum deviation of 2.6. This computer model can be used to improve exposure limits or for a quantitative risk analysis of a given exposure of the skin.
国际非电离辐射防护委员会(ICNIRP)、国际电工委员会(IEC)60825-1标准以及美国国家标准学会(ANSI)Z136.1标准中用于保护皮肤的暴露和发射限值是基于有限数量的研究制定的。为了拓宽数据库,开发了一种计算机模型,用于预测400纳米至20微米波长范围内的损伤阈值,并通过与488纳米至10.6微米波长范围以及8微秒至630秒暴露持续时间内的所有适用实验阈值数据(ED)进行比较来验证。模型预测结果与数据相比表现良好,计算机预测值与ED的平均比值为1.01(标准偏差±46%),最大偏差为2.6。该计算机模型可用于改进暴露限值或对皮肤的给定暴露进行定量风险分析。