Boothe W A, Lee D A, Panek W C, Pettit T H
Jules Stein Eye Institute, UCLA Medical Center 90024.
Arch Ophthalmol. 1988 Sep;106(9):1214-7. doi: 10.1001/archopht.1988.01060140374035.
A new miniaturized, digital, electronic tonometer, the Tono-Pen (Intermedics, Pasadena, Calif), was evaluated for accuracy and reproducibility of results in the laboratory using enucleated human eyes with an indwelling pressure transducer (manometric study) and clinically using 100 eyes in 50 patients with normal corneas. Reproducibility was good in both the manometric study (SD = 1.09) and in the clinical study (SD = 1.21). Reproducibility of results was similar to that of the MacKay-Marg tonometer and the pneumotonometer in the manometric study, but less than that of the Goldmann applanation tonometer and the pneumotonometer in the clinical study. In the manometric study, the Tono-Pen had less mean error at high pressure ranges than either the MacKay-Marg tonometer or the pneumotonometer. In the clinical study, no statistically significant difference was found between the Tono-Pen and the applanation tonometer at pressures of 10 to 35 mm Hg (the extent of the data spread). In addition to being reliable and accurate, this miniaturized tonometer is portable and easy to use.
一种新型的小型数字式电子眼压计——托诺笔眼压计(Intermedics公司,加利福尼亚州帕萨迪纳),在实验室中,利用植入压力传感器的摘除人眼(压力测量研究)对其结果的准确性和可重复性进行了评估;在临床中,对50例角膜正常患者的100只眼睛进行了评估。在压力测量研究(标准差=1.09)和临床研究(标准差=1.21)中,可重复性都很好。在压力测量研究中,该眼压计结果的可重复性与麦凯-马格眼压计和气动眼压计相似,但在临床研究中,低于戈德曼压平眼压计和气动眼压计。在压力测量研究中,托诺笔眼压计在高压范围内的平均误差比麦凯-马格眼压计或气动眼压计都小。在临床研究中,在10至35毫米汞柱压力范围(数据分布范围)内,托诺笔眼压计与压平眼压计之间未发现统计学上的显著差异。除了可靠和准确外,这种小型眼压计便于携带且易于使用。