两种压平眼压计在猫中的评估。

Evaluation of two applanation tonometers in cats.

作者信息

Miller P E, Pickett J P, Majors L J, Kurzman I D

机构信息

Department of Surgical Science, School of Veterinary Medicine, University of Wisconsin-Madison 53706.

出版信息

Am J Vet Res. 1991 Nov;52(11):1917-21.

PMID:1785739
Abstract

Comparisons of the MacKay-Marg and Tono-Pen applanation tonometers in open and closed in vitro systems were made for the eyes of cats. Both instruments significantly underestimated intraocular pressure (IOP) vs direct manometry (P less than 0.001), but in readily predictable manner, with high coefficients of determination (r2 = 0.99). For tonometer 1 (MacKay-Marg), calculated actual IOP = 1.36 x (MacKay-Marg measurement) - 1.67 mm of Hg; and for tonometer 2 (Tono-Pen), calculated actual IOP = 1.37 x (Tono-Pen measurement) + 0.8 mm of HG, using measurements from 11 enucleated eyes. In vivo comparisons were initially made in 81 clinically normal eyes (n = 41 cats) by applying the Tono-Pen first followed by the MacKay-Marg. Compared with the MacKay-Marg, the Tono-Pen significantly (P less than 0.001) underestimated IOP in these cats. When the order of tonometer applanation was subsequently reversed in 73 clinically normal eyes (n = 37 cats) the Tono-Pen again significantly (P less than 0.001) underestimated IOP, compared with the MacKay-Marg. Alterations in tonometer order did not result in significant differences in measured IOP for the MacKay-Marg when compared with itself, but Tono-Pen measurements were significantly (P less than 0.05) less when its use followed, rather than preceded, that of the MacKay-Marg. Mean (+/- SD) IOP in clinically normal cats when each tonometer was used first was 22.6 +/- 4.0 mm of Hg (range, 14 to 32 mm of Hg) for the MacKay-Marg and 19.7 +/- 5.6 mm of Hg (9 to 31 mm of Hg) for the Tono-Pen.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

在开放和封闭的体外系统中,对猫的眼睛使用麦凯-马格(MacKay-Marg)眼压计和托诺笔(Tono-Pen)眼压计进行了比较。与直接测压法相比,两种仪器均显著低估了眼压(IOP)(P<0.001),但方式易于预测,决定系数较高(r2 = 0.99)。对于眼压计1(麦凯-马格),使用11只摘除眼球的测量数据,计算得出的实际眼压=1.36×(麦凯-马格测量值)-1.67毫米汞柱;对于眼压计2(托诺笔),计算得出的实际眼压=1.37×(托诺笔测量值)+0.8毫米汞柱。最初在81只临床正常的眼睛(n = 41只猫)中进行了体内比较,先使用托诺笔,然后使用麦凯-马格。与麦凯-马格相比,托诺笔在这些猫中显著(P<0.001)低估了眼压。随后在73只临床正常的眼睛(n = 37只猫)中颠倒眼压计压平顺序,与麦凯-马格相比,托诺笔再次显著(P<0.001)低估了眼压。与自身相比,眼压计顺序的改变并未导致麦凯-马格测量的眼压出现显著差异,但当托诺笔在麦凯-马格之后而非之前使用时,其测量值显著(P<0.05)更低。当每种眼压计首先使用时,临床正常猫的平均(±标准差)眼压,麦凯-马格为22.6±4.0毫米汞柱(范围为14至32毫米汞柱),托诺笔为19.7±5.6毫米汞柱(9至31毫米汞柱)。(摘要截断于250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验