The accuracy of intraocular pressure measurement with the minified Goldmann applanation tonometer, the pneumatonometer, and the Tono-Pen tonometer were compared in post-mortem human eyes which had undergone penetrating keratoplasty. Enucleated post-mortem human eyes underwent same sized (7.75 mm) or 0.5 mm oversized (8.25 mm) autologous penetrating keratoplasty. Intraocular pressure was then set and measured manometrically while being determined successively with each tonometer over the range of 0-65 mm Hg. Linear regression analysis comparing tonometric and manometric readings showed: (1) minified Goldmann applanation tonometer-slope 0.985 and 0.944, intercept 1.64 and 2.55 mm Hg, correlation coefficient 0.99 and 0.99 in same sized and oversized grafted eyes respectively; (2) pneumatonometer-slope 1.008 and 0.990, intercept 3.37 and 3.69 mm Hg, correlation coefficient 0.99 and 0.98; (3) Tono-Pen-slope 1.061 and 1.002, intercept 5.01 and 4.06 mm Hg, correlation coefficient 0.97 and 0.98. We concluded that the minified Goldmann applanation tonometer is as accurate or more accurate than the pneumatonometer and the Tono-Pen in post-mortem post-keratoplasty human eyes, and may be an economical, convenient alternative to the latter two instruments in clinical practice.
在接受穿透性角膜移植术的人死后眼球中,比较了使用小型戈德曼压平眼压计、气动眼压计和托诺笔眼压计测量眼压的准确性。摘除的人死后眼球接受了相同尺寸(7.75毫米)或0.5毫米超大尺寸(8.25毫米)的自体穿透性角膜移植术。然后设定眼压并进行压力测量,同时在0至65毫米汞柱的范围内依次用每种眼压计进行测量。比较眼压计读数和压力测量读数的线性回归分析显示:(1)小型戈德曼压平眼压计——在相同尺寸和超大尺寸移植眼中,斜率分别为0.985和0.944,截距分别为1.64和2.55毫米汞柱,相关系数分别为0.99和0.99;(2)气动眼压计——斜率分别为1.008和0.990,截距分别为3.37和3.69毫米汞柱,相关系数分别为0.99和0.98;(3)托诺笔眼压计——斜率分别为1.061和1.002,截距分别为5.01和4.06毫米汞柱,相关系数分别为0.97和0.98。我们得出结论,在人死后角膜移植后的眼球中,小型戈德曼压平眼压计与气动眼压计和托诺笔眼压计一样准确或更准确,在临床实践中可能是后两种仪器的一种经济、方便的替代方案。