Department of Ophthalmology, The David J. Apple International Laboratory for Ocular Pathology, University of Heidelberg, Heidelberg, Germany.
PLoS One. 2021 Jul 19;16(7):e0254901. doi: 10.1371/journal.pone.0254901. eCollection 2021.
Experimental study to measure the intraocular lens (IOL) injection time and injection speed at different intraocular pressure (IOP) settings when using the AutonoMe® injector.
In this experimental study, following phacoemulsification in porcine cadaver eyes, a trocar was inserted at pars plana with a connected infusion and IOPs of 20, 50 and 80 mmHg were generated by altering the infusion height. Twelve CO2 gas-driven injectors were used to implant an IOL via a corneal incision of 2.2 mm. For each IOP setting, the duration of the IOL injection and the injection speed was measured by analyzing a video recording of the procedure.
The mean ±SD injection time (seconds) was 4.47±0.50 at 20 mmHg, 4.98±0.55 at 50 mmHg and 5.47±0.20 at 80 mmHg. The mean ±SD injection speed (millimeters per seconds) was 1.36±0.15 at 20 mmHg, 1.22±0.14 at 50 mmHg and 1.10±0.04 at 80 mmHg. There was a significant (p<0.05) difference between the 20 and 80 mmHg groups in mean injection duration and injection speed.
The CO2 gas driven injector allows a safe IOL injection even at elevated IOP. Although the implantation time is slightly extended at higher IOPs, this does not seem to be clinically relevant. No IOL damage was observed at these pressure settings.
使用 AutonoMe®注射器,在不同眼压(IOP)设置下测量眼内人工晶状体(IOL)注射时间和注射速度的实验研究。
在这项实验研究中,在猪尸体眼球的超声乳化后,通过插入巩膜突并连接输液,通过改变输液高度产生 20、50 和 80mmHg 的 IOP。使用 12 个 CO2 气体驱动注射器通过 2.2mm 的角膜切口植入 IOL。对于每个 IOP 设置,通过分析手术过程的视频记录来测量 IOL 注射的持续时间和注射速度。
平均±SD 注射时间(秒)在 20mmHg 时为 4.47±0.50,在 50mmHg 时为 4.98±0.55,在 80mmHg 时为 5.47±0.20。平均±SD 注射速度(毫米/秒)在 20mmHg 时为 1.36±0.15,在 50mmHg 时为 1.22±0.14,在 80mmHg 时为 1.10±0.04。在平均注射持续时间和注射速度方面,20mmHg 和 80mmHg 组之间存在显著差异(p<0.05)。
即使在高眼压下,CO2 气体驱动注射器也允许安全地注射 IOL。尽管在较高的 IOP 下植入时间略有延长,但这似乎在临床上并不相关。在这些压力设置下没有观察到 IOL 损坏。