Determination of a cone-beam CT low-dose protocol for root fracture diagnosis in non-endodontically treated anterior maxillary teeth.

机构信息

College of Dentistry, Hawler Medical University, Erbil, Iraq.

Division of Dentistry, School of Medical Sciences, Faculty of Biology, Medicine andHealth, University of Manchester, Manchester Academic Health Science Centre, Manchester, UK.

出版信息

Dentomaxillofac Radiol. 2022 Feb 1;51(2):20210138. doi: 10.1259/dmfr.20210138. Epub 2021 Sep 8.

Abstract

OBJECTIVES

The aim of this study was to determine a "low-dose protocol" which provides acceptable diagnostic accuracy for detection of root fractures in unrestored anterior maxillary teeth, using an model.

METHODS

48 maxillary anterior teeth, half with horizontal or oblique root fractures, were imaged using CBCT in an anthropomorphic model. Nine X-ray exposure combinations were used, including the manufacturer's standard ("reference") exposure and high-resolution settings ("HiRes"), by varying kV, exposure time, and rotation angle. Measurements of Dose Area Product (DAP) were recorded. Five dental radiologists assessed the scans for root fractures and judged image quality. Parameters of diagnostic accuracy were calculated, including area under the Receiver Operating Characteristic curve (A). Objective measures of image quality were made at the same exposure combinations using an image quality phantom.

RESULTS

Although there was a significant linear relationship between DAP and mean A, only the lowest DAP exposure combination had a mean A significantly different to the reference exposure. There was no significant effect on other diagnostic accuracy parameters when using HiRes compared with the reference exposure. There was a significant positive relationship between DAP and contrast resolution. HiRes did not significantly improve contrast resolution and made a small improvement to spatial resolution.

CONCLUSIONS

Scope existed for radiation dose reduction compared with the manufacturer's guidance. There was no improvement in diagnostic accuracy using HiRes settings. A cautious recommendation for this CBCT machine is that it is possible to achieve a dose reduction of about 20% compared with the reference exposure parameters.

摘要

目的

本研究旨在通过体模确定一种“低剂量方案”,以提供可接受的诊断准确性,用于检测未修复的上颌前牙的根折。

方法

使用 CBCT 对 48 颗上颌前牙(其中一半有水平或斜形根折)进行成像,这些牙位于人体模型中。使用了 9 种 X 射线曝光组合,包括制造商的标准(“参考”)曝光和高分辨率设置(“HiRes”),通过改变 kV、曝光时间和旋转角度来实现。记录剂量面积乘积(DAP)的测量值。5 名牙科放射科医生对扫描结果进行根折评估,并判断图像质量。计算了包括受试者工作特征曲线下面积(A)在内的诊断准确性参数。使用图像质量体模在相同的曝光组合下进行了客观的图像质量测量。

结果

尽管 DAP 与平均 A 之间存在显著的线性关系,但只有最低 DAP 曝光组合的平均 A 与参考曝光显著不同。与参考曝光相比,使用 HiRes 时对其他诊断准确性参数没有显著影响。DAP 与对比度分辨率之间存在显著的正相关关系。HiRes 并没有显著提高对比度分辨率,只是对空间分辨率略有改善。

结论

与制造商的指南相比,存在降低辐射剂量的空间。使用 HiRes 设置并不能提高诊断准确性。对于这种 CBCT 机,谨慎的建议是,与参考曝光参数相比,有可能降低约 20%的剂量。