Elbardesy Hany, Yadav Himanshu, Rabea Mohamed, Guerin Shane, Harty James
Department of Trauma and Orthopaedic, Cork University Hospital, Wilton, Cork T12DFK4, Ireland.
Faculty of Pharmacy, Mansoura University, Mansoura 32216, Egypt.
SICOT J. 2021;7:47. doi: 10.1051/sicotj/2021045. Epub 2021 Sep 10.
Debate encompasses the use of Vitamin E Polyethylene or conventional Polyethylene liner in primary hip arthroplasty. Does the Inclusion of Vitamin E in PE give adequate protection from oxidation and maintains lower rates of wear?
We performed this study following the Preferred Reporting Items for Systematic Reviews and Meta-analyses Statement (PRISMA) and the Cochrane Handbook for systematic reviews and meta-analysis. Studies were included from any region, written in any language. We had only the randomised control trials comparing the femoral head penetration between Vitamin-E diffused highly cross-linked polyethylene (VEPE) liner and conventional liners in primary total hip arthroplasty.
We included 10 studies in this meta-analysis. We conducted them using Review Manager V.5.0. We computed the risk ratio to measure the treatment effect, considering the heterogeneity. We used Random-effect models. VEPE had insignificant marginal advantages for FHP within three months post-operative. Additionally, VEPE showed significantly less FHP after two and five years. After one year, it showed significantly less FHP with the VEPE group versus the UHMWPE cohort and a non-significant difference between the VEPE and XLPE group.
In terms of FHP, this metanalysis shows less FHP for the VEPE than conventional PE. A longer follow-up period is required to evaluate whether the oxidation protection gained by Vitamin E results in lower wear rates, less osteolysis, and aseptic loosening compared to the conventional PE in the long term.
关于在初次髋关节置换术中使用维生素E聚乙烯内衬还是传统聚乙烯内衬存在争议。在聚乙烯中添加维生素E是否能提供足够的抗氧化保护并保持较低的磨损率?
我们按照系统评价和Meta分析的首选报告项目声明(PRISMA)以及Cochrane系统评价和Meta分析手册进行了这项研究。纳入来自任何地区、用任何语言撰写的研究。我们仅纳入了比较初次全髋关节置换术中维生素E扩散高交联聚乙烯(VEPE)内衬与传统内衬之间股骨头穿透情况的随机对照试验。
我们在这项Meta分析中纳入了10项研究。使用Review Manager V.5.0进行分析。考虑到异质性,我们计算风险比来衡量治疗效果。我们使用随机效应模型。术后三个月内,VEPE在股骨头穿透方面的边际优势不显著。此外,在术后两年和五年时,VEPE的股骨头穿透明显更少。术后一年时,与超高分子量聚乙烯(UHMWPE)队列相比,VEPE组的股骨头穿透明显更少,且VEPE组与交联聚乙烯(XLPE)组之间无显著差异。
就股骨头穿透而言,这项Meta分析表明VEPE的股骨头穿透比传统聚乙烯少。需要更长的随访期来评估与传统聚乙烯相比,维生素E提供的抗氧化保护从长期来看是否能导致更低的磨损率、更少的骨溶解和无菌性松动。