剂量计校准因子(N)在20年期间的变化。

Variation in Dosimeter Calibration Factor (N) Over a Period of 20 Years.

作者信息

Semwal Manoj Kumar, Gupta Harishchandra, Singhal Sakshi, Kumar Ashok, Kaushik Aruna

机构信息

Department of Radiation Oncology, Army Hospital Research and Referral, Delhi, India.

Division of Clinical PET, Institute of Nuclear Medicine and Allied Sciences, Delhi, India.

出版信息

J Med Phys. 2021 Apr-Jun;46(2):125-127. doi: 10.4103/jmp.JMP_42_21. Epub 2021 Aug 7.

Abstract

Long delays in renewal of calibration of secondary standards radiation dosimeters in radiation oncology centers due to the COVID19 pandemic have aroused concerns regarding accuracy in dose delivery to patients. The concerns are mainly due to the uncertainty in the absorbed dose to water calibration factor (N) over a period of time. In this study, the Nfactor for two ion chambers, thimble type (Farmer) and parallel plate type (Markus), used in most of the radiotherapy centers, were retrospectively reviewed for 20 years. The calibration on all occasions except once was carried out at the Secondary Standards Dosimetry Laboratory, Bhabha Atomic Research Centre, Mumbai. The change in the Nfactor over this period was < ±3%. We, therefore, believe that a dosimeter with no history of repairs showed reasonable stability in the Nfactor over a long period.

摘要

由于新冠疫情,放射肿瘤中心二级标准辐射剂量计校准更新的长时间延迟引发了对患者剂量输送准确性的担忧。这些担忧主要源于一段时间内水吸收剂量校准因子(N)的不确定性。在本研究中,回顾性分析了大多数放疗中心使用的两种电离室(指形电离室(Farmer型)和平行板电离室(Markus型))20年的N因子。除一次外,所有校准均在孟买巴巴原子研究中心的二级标准剂量学实验室进行。在此期间,N因子的变化<±3%。因此,我们认为,一台无维修历史的剂量计在很长一段时间内N因子表现出合理的稳定性。

相似文献

1
Variation in Dosimeter Calibration Factor (N) Over a Period of 20 Years.剂量计校准因子(N)在20年期间的变化。
J Med Phys. 2021 Apr-Jun;46(2):125-127. doi: 10.4103/jmp.JMP_42_21. Epub 2021 Aug 7.