Suppr超能文献

自动化细胞计数与体液分析参考方法的性能评估比较。

Performance evaluation of automated cell counts compared with reference methods for body fluid analysis.

机构信息

Clinical Chemistry Laboratory, Papa Giovanni XXIII Hospital, Bergamo, Italy.

Regional Reference Center for the Quality of Laboratory Medicine Services, Milan, Italy.

出版信息

Diagnosis (Berl). 2021 Oct 1;9(3):369-378. doi: 10.1515/dx-2021-0088. eCollection 2022 Aug 1.

Abstract

OBJECTIVES

Cellular analysis of body fluids (BFs) can assist clinicians for the diagnosis of many medical conditions. The aim of this work is the evaluation of the analytical performance of the UF-5000 body fluid mode (UF-BF) analyzer compared to the gold standard method (optical microscopy, OM) and to XN-1000 (XN-BF), another analyzer produced by the same manufacturer (Sysmex) and with a similar technology for BF analysis.

METHODS

One hundred BF samples collected in KEDTA tubes were analyzed by UF-BF, XN-BF and OM. The agreement was evaluated using Passing and Bablok regression and Bland-Altman plot analysis. The receiver operating characteristic (ROC) curves were selected for evaluating the diagnostic agreement between OM classification and UF-BF parameters.

RESULTS

Comparison between UF-BF and OM, in all BF types, showed Passing and Bablok's slope comprised between 0.99 (polymorphonuclear cells count, PMN-BF) and 1.39 (mononuclear cells count, MN-BF), the intercepts ranged between 26.47 (PMN-BF parameter) and 226.80 (white blood cell count). Bland-Altman bias was comprised between 7.3% (total cell count, TC-BF) and 52.9% (MN-BF). Comparison between UF-BF and XN-BF in all BF showed slopes ranged between 1.07 (TC-BF and PMN-BF) and 1.16 (MN-BF), intercepts ranged between 8.30 (PMN) and 64.78 (WBC-BF). Bland-Altman bias ranged between 5.8 (TC-BF) and 21.1% (MN-BF). The ROC curve analysis showed an area under the curve ranged between 0.9664 and 1.000.

CONCLUSIONS

UF-BF shows very good performance for the differential counts of cells in ascitic, pleural and synovial fluids and therefore it is useful to screen and count cells in this type of BF.

摘要

目的

体液(BFs)的细胞分析有助于临床医生诊断许多医学病症。本工作的目的是评估 UF-5000 体液模式(UF-BF)分析仪与金标准方法(光学显微镜,OM)以及另一种由同一制造商(Sysmex)生产的类似 BF 分析技术的 XN-1000(XN-BF)分析仪的分析性能。

方法

使用 UF-BF、XN-BF 和 OM 分析了收集在 KEDTA 管中的 100 个 BF 样本。使用 Passing 和 Bablok 回归以及 Bland-Altman 图分析评估了一致性。为了评估 OM 分类与 UF-BF 参数之间的诊断一致性,选择了接收者操作特征(ROC)曲线。

结果

在所有 BF 类型中,UF-BF 与 OM 的比较显示 Passing 和 Bablok 的斜率在 0.99(多形核细胞计数,PMN-BF)和 1.39(单核细胞计数,MN-BF)之间,截距范围在 26.47(PMN-BF 参数)和 226.80(白细胞计数)之间。Bland-Altman 偏差在 7.3%(总细胞计数,TC-BF)和 52.9%(MN-BF)之间。在所有 BF 中,UF-BF 与 XN-BF 的比较显示斜率在 1.07(TC-BF 和 PMN-BF)和 1.16(MN-BF)之间,截距在 8.30(PMN)和 64.78(WBC-BF)之间。Bland-Altman 偏差在 5.8(TC-BF)和 21.1%(MN-BF)之间。ROC 曲线分析显示曲线下面积在 0.9664 到 1.000 之间。

结论

UF-BF 对腹水、胸腔积液和滑液中的细胞差异计数表现出非常好的性能,因此它有助于筛选和计数此类 BF 中的细胞。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验