XN 体液(BF)模块对脑脊液细胞计数和分类的双部位评估诊断性能。

Two-site evaluation of the diagnostic performance of the Sysmex XN Body Fluid (BF) module for cell count and differential in Cerebrospinal Fluid.

机构信息

Clinical Chemistry Laboratory, Papa Giovanni XXIII Hospital, Bergamo, Italy.

Clinical Chemistry Laboratory Hospital Careggi Firenze, University of Firenze, Firenze, Italy.

出版信息

Int J Lab Hematol. 2018 Feb;40(1):26-33. doi: 10.1111/ijlh.12723. Epub 2017 Sep 3.

Abstract

INTRODUCTION

Cellular analysis in cerebrospinal fluid (CSF) provides important diagnostic information in many pathological settings. The aim of this two-site study was to evaluate the Sysmex XN Body Fluid mode (XN-BF) for cell analysis of CSF compared to light microscopy (LM).

METHODS

Two hundred and seven consecutive CSF samples were analyzed in parallel with XN-BF and LM. The study also included the estimation of the limit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ), carry-over and linearity of XN-BF module.

RESULTS

LoQ of white blood cells (WBC) was 3×10  cells/L; linearity was good and carry-over negligible. XN-BF parameters were compared to LM for the following cell classes: total cells, WBC, polymorphonuclear (PMN), and mononuclear (MN) cells. The bias ranged from 1.3 to 15.2×10  cells/L. The receiver operating characteristics curve analysis for WBC showed an area under the curve of 0.98, and the global diagnostic agreement was 95% at a cutoff of 5×10  cells/L.

CONCLUSIONS

XN-BF provides rapid and accurate counts in clinically relevant ranges of CSF values, thus providing a valuable alternative to conventional LM analysis. However, microscopic review remains advisable in samples with abnormal cell counts or high fluorescent (HF-BF) cell parameter exceeding 5×10  cells/L.

摘要

简介

脑脊液(CSF)中的细胞分析在许多病理情况下提供了重要的诊断信息。本两项研究的目的是评估 Sysmex XN 体液模式(XN-BF)在 CSF 细胞分析方面与光学显微镜(LM)相比的性能。

方法

207 例连续 CSF 样本与 XN-BF 和 LM 平行分析。该研究还包括对空白限(LoB)、检测限(LoD)、定量限(LoQ)、交叉污染和 XN-BF 模块线性度的估计。

结果

白细胞(WBC)的 LoQ 为 3×10 个细胞/L;线性度良好,交叉污染可忽略不计。XN-BF 参数与 LM 比较以下细胞类别:总细胞、WBC、多形核(PMN)和单核(MN)细胞。偏差范围为 1.3 至 15.2×10 个细胞/L。WBC 的受试者工作特征曲线分析显示曲线下面积为 0.98,在 5×10 个细胞/L 的截定点处,总体诊断一致性为 95%。

结论

XN-BF 可在临床相关的 CSF 值范围内提供快速准确的计数,因此是传统 LM 分析的有价值的替代方法。然而,在细胞计数异常或荧光(HF-BF)细胞参数超过 5×10 个细胞/L 的样本中,仍建议进行显微镜检查。