Cellular analysis in cerebrospinal fluid (CSF) provides important diagnostic information in many pathological settings. The aim of this two-site study was to evaluate the Sysmex XN Body Fluid mode (XN-BF) for cell analysis of CSF compared to light microscopy (LM).
Two hundred and seven consecutive CSF samples were analyzed in parallel with XN-BF and LM. The study also included the estimation of the limit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ), carry-over and linearity of XN-BF module.
LoQ of white blood cells (WBC) was 3×10 cells/L; linearity was good and carry-over negligible. XN-BF parameters were compared to LM for the following cell classes: total cells, WBC, polymorphonuclear (PMN), and mononuclear (MN) cells. The bias ranged from 1.3 to 15.2×10 cells/L. The receiver operating characteristics curve analysis for WBC showed an area under the curve of 0.98, and the global diagnostic agreement was 95% at a cutoff of 5×10 cells/L.
XN-BF provides rapid and accurate counts in clinically relevant ranges of CSF values, thus providing a valuable alternative to conventional LM analysis. However, microscopic review remains advisable in samples with abnormal cell counts or high fluorescent (HF-BF) cell parameter exceeding 5×10 cells/L.
脑脊液(CSF)中的细胞分析在许多病理情况下提供了重要的诊断信息。本两项研究的目的是评估 Sysmex XN 体液模式(XN-BF)在 CSF 细胞分析方面与光学显微镜(LM)相比的性能。
207 例连续 CSF 样本与 XN-BF 和 LM 平行分析。该研究还包括对空白限(LoB)、检测限(LoD)、定量限(LoQ)、交叉污染和 XN-BF 模块线性度的估计。
白细胞(WBC)的 LoQ 为 3×10 个细胞/L;线性度良好,交叉污染可忽略不计。XN-BF 参数与 LM 比较以下细胞类别:总细胞、WBC、多形核(PMN)和单核(MN)细胞。偏差范围为 1.3 至 15.2×10 个细胞/L。WBC 的受试者工作特征曲线分析显示曲线下面积为 0.98,在 5×10 个细胞/L 的截定点处,总体诊断一致性为 95%。
XN-BF 可在临床相关的 CSF 值范围内提供快速准确的计数,因此是传统 LM 分析的有价值的替代方法。然而,在细胞计数异常或荧光(HF-BF)细胞参数超过 5×10 个细胞/L 的样本中,仍建议进行显微镜检查。