体液模式下腹水、胸腔积液、滑液和脑脊液中高荧光细胞自动计数缺乏协调性。

Lack of harmonization in high fluorescent cell automated counts with body fluids mode in ascitic, pleural, synovial, and cerebrospinal fluids.

机构信息

Clinical Chemistry Laboratory, Hospital Papa Giovanni XXIII, Bergamo, Italy.

Synlab Italia Srl, Castenedolo, Italy.

出版信息

Int J Lab Hematol. 2019 Apr;41(2):277-286. doi: 10.1111/ijlh.12968. Epub 2019 Feb 13.

Abstract

INTRODUCTION

Cellular analysis in body fluids (BFs) provides important diagnostic information in various pathological settings. This study was hence aimed at comparing automated cell count obtained with Mindray BC-6800 (BC-BF) vs Sysmex XN-series (XN-BF) and evaluating other quantitative and qualitative information provided by these analyzers in ascitic (AF), pleural (PF), synovial (SF), and cerebrospinal (CSF) fluids.

METHODS

Three hundred and fifty-one samples (99 AFs, 45 PFs, 75 SFs, and 132 CSFs) were analyzed in parallel with BC-BF, XN-BF, and optical microscopy (OM). The study also included the assessment of diagnostic agreement among BC-BF, XN-BF, and OM.

RESULTS

The comparison of BC-BF vs XN-BF yielded slopes of Passing and Bablok regression always comprised between 0.9 and 1.0 except for EO-BF and HF-BF, whilst the intercepts ranged from -0.4 for MN-BF and 12.0 for PMN-BF. The bias was comprised between -103.3% and 67.1% for HF-BF and EO-BF, respectively. A significant bias was found for TC-BF, WBC-BF, HF-BF (negative bias) and for PMN-BF and EO-BF (positive bias). The agreement (Cohen's kappa) between XN-BF and BC-BF was always ≥0.7, ranging between 0.87 in CSFs and 0.94 in AFs, and that with OM was similar (ie, 0.85 and 0.96).

CONCLUSION

The cytometric analysis of BF samples using BC-BF and XN-BF is clinically favorable when appropriately combined with OM. Quantitative and qualitative parameters displayed optimal agreement, whilst instrument-specific cut-offs should be defined and implemented for HF-BF and EO-BF. Further efforts should be made for achieving better harmonization in cytometric analysis of BF samples.

摘要

简介

体液(BF)中的细胞分析在各种病理情况下提供重要的诊断信息。因此,本研究旨在比较 Mindray BC-6800(BC-BF)与 Sysmex XN 系列(XN-BF)自动细胞计数,并评估这些分析仪在腹水(AF)、胸腔积液(PF)、滑液(SF)和脑脊液(CSF)中提供的其他定量和定性信息。

方法

对 351 个样本(99 个 AFs、45 个 PF、75 个 SF 和 132 个 CSF)进行了平行分析,包括 BC-BF、XN-BF 和光学显微镜(OM)。本研究还包括评估 BC-BF、XN-BF 和 OM 之间的诊断一致性。

结果

BC-BF 与 XN-BF 的比较结果得出的斜率总是在 0.9 到 1.0 之间,除了 EO-BF 和 HF-BF 之外,而截距范围从 MN-BF 的-0.4 到 PMN-BF 的 12.0。HF-BF 和 EO-BF 的偏差分别为-103.3%和 67.1%。TC-BF、WBC-BF、HF-BF(负偏差)和 PMN-BF 和 EO-BF(正偏差)的偏差均显著。XN-BF 与 BC-BF 之间的一致性(Cohen's kappa)始终≥0.7,范围从 CSF 中的 0.87 到 AF 中的 0.94,与 OM 的一致性相似(即 0.85 和 0.96)。

结论

当与 OM 适当结合时,BC-BF 和 XN-BF 对 BF 样本的细胞计量分析在临床上是有利的。定量和定性参数显示出最佳的一致性,而 HF-BF 和 EO-BF 应定义和实施仪器特定的截止值。应进一步努力实现 BF 样本细胞计量分析的更好协调。