Shi Mengying, Chuang Cynthia F, Kovalchuk Nataliya, Bush Karl, Zaks Daniel, Xing Lei, Surucu Murat, Han Bin
Department of Radiation Oncology, Stanford University, Stanford, California, USA.
RefleXion Medical, Hayward, California, USA.
Med Phys. 2021 Nov;48(11):7450-7460. doi: 10.1002/mp.15273. Epub 2021 Oct 22.
The RefleXion™ X1 is a novel radiotherapy system that is designed for image-guided radiotherapy, and eventually, biology-guided radiotherapy (BgRT). BgRT is a treatment paradigm that tracks tumor motion using real-time positron emission signals. This study reports the small-field measurement results and the validation of a Monte Carlo (MC) model of the first clinical RefleXion unit.
The RefleXion linear accelerator (linac) produces a 6 MV flattening filter free (FFF) photon beam and consists of a binary multileaf collimator (MLC) system with 64 leaves and two pairs of y-jaws. The maximum clinical field size achievable is 400 × 20 mm . The y-jaws provide either a 10 or 20 mm opening at source-to-axis distance (SAD) of 850 mm. The width of each MLC leaf at SAD is 6.25 mm. Percentage depth doses (PDDs) and relative beam profiles were acquired using an Edge diode detector in a water tank for field sizes from 12.5 × 10 to 100 × 20 mm . Beam profiles were also measured using films. Output factors of fields ranging from 6.25 × 10 to 100 × 20 mm were measured using W2 scintillator detector, Edge detector, and films. Output correction factors k of the Edge detector for RefleXion were calculated. An MC model of the linac including pre-MLC beam sources and detailed structures of MLC and lower y-jaws was validated against the measurements. Simulation codes BEAMnrc and GATE were utilized.
The diode measured PDD at 10 cm depth (PDD10) increases from 53.6% to 56.9% as the field opens from 12.5 × 10 to 100 × 20 mm . The W2-measured output factor increases from 0.706 to 1 as the field opens from 6.25 × 10 to 100 × 20 mm (reference field size). The output factors acquired by diode and film differ from the W2 results by 1.65% (std = 1.49%) and 2.09% (std = 1.41%) on average, respectively. The profile penumbra and full-width half-maximum (FWHM) measured by diode agree well with the film results with a deviation of 0.60 mm and 0.73% on average, respectively. The averaged beam profile consistency calculated between the diode- and film-measured profiles among different depths is within 1.72%. By taking the W2 measurements as the ground truth, the output correction factors k for Edge detector ranging from 0.958 to 1 were reported. For the MC model validation, the simulated PDD10 agreed within 0.6% to the diode measurement. The MC-simulated output factor differed from the W2 results by 2.3% on average (std = 3.7%), while the MC simulated beam penumbra differed from the diode results by 0.67 mm on average (std = 0.42 mm). The MC FWHM agreed with the diode results to within 1.40% on average. The averaged beam profile consistency calculated between the diode and MC profiles among different depths is less than 1.29%.
This study represents the first small-field dosimetry of a clinical RefleXion system. A complete and accurate MC model of the RefleXion linac has been validated.
RefleXion™ X1是一款新型放射治疗系统,专为图像引导放射治疗设计,最终目标是实现生物引导放射治疗(BgRT)。BgRT是一种利用实时正电子发射信号追踪肿瘤运动的治疗模式。本研究报告了首个临床RefleXion装置的小射野测量结果以及蒙特卡罗(MC)模型的验证情况。
RefleXion直线加速器(直线加速器)产生6兆伏无均整器(FFF)光子束,由一个具有64个叶片的二元多叶准直器(MLC)系统和两对y形准直器组成。可实现的最大临床射野尺寸为400×20毫米。y形准直器在源轴距(SAD)为850毫米时提供10或20毫米的开口。每个MLC叶片在SAD处的宽度为6.25毫米。使用水箱中的Edge二极管探测器获取了射野尺寸从12.5××10到100×20毫米的百分深度剂量(PDD)和相对射束轮廓。还使用胶片测量了射束轮廓。使用W2闪烁探测器、Edge探测器和胶片测量了射野尺寸从6.25×10到100×20毫米的输出因子。计算了RefleXion的Edge探测器的输出校正因子k。针对测量结果对包括MLC前射束源以及MLC和下部y形准直器详细结构的直线加速器MC模型进行了验证。使用了模拟代码BEAMnrc和GATE。
随着射野从12.5×10扩大到100×20毫米,二极管在10厘米深度处测量的PDD(PDD10)从53.6%增加到56.9%。随着射野从6.25×10扩大到100×20毫米(参考射野尺寸),W2测量的输出因子从0.706增加到1。二极管和胶片获取的输出因子与W2结果的平均差异分别为1.65%(标准差=1.49%)和2.09%(标准差=1.41%)。二极管测量的轮廓半值宽度和半高宽(FWHM)与胶片结果吻合良好,平均偏差分别为0.60毫米和0.73%。不同深度之间二极管和胶片测量轮廓之间计算的平均射束轮廓一致性在1.72%以内。以W2测量结果为基准,报告了Edge探测器的输出校正因子k范围为0.958至1。对于MC模型验证,模拟的PDD10与二极管测量结果的偏差在0.6%以内。MC模拟的输出因子与W2结果的平均差异为2.3%(标准差=3.7%),而MC模拟的射束半值宽度与二极管结果的平均差异为0.67毫米(标准差=0.42毫米)。MC FWHM与二极管结果的平均吻合度在±1.40%以内。不同深度之间二极管和MC轮廓之间计算的平均射束轮廓一致性小于1.29%。
本研究代表了首个临床RefleXion系统的小射野剂量学研究。已验证了RefleXion直线加速器完整且准确的MC模型。