秘鲁常见X射线检查程序的诊断参考水平。

Diagnostic Reference Levels for Common X-ray Procedures in Peru.

作者信息

Portocarrero Bonifaz Andres, Camarena Rodriguez Caterina S, Palma Esparza Ricardo

机构信息

Radiation Oncology, University of Louisville, Louisville, USA.

Physics, Pontificia Universidad Catolica del Peru, Lima, PER.

出版信息

Cureus. 2021 Oct 7;13(10):e18566. doi: 10.7759/cureus.18566. eCollection 2021 Oct.

Abstract

Diagnostic reference levels (DRLs) for X-ray procedures have been established in many countries since 1996. In Peru, data from the literature are used as guidelines as DRLs research is limited. The objective of this study is to analyze the parameters and variables which are used in radiological techniques such as kV, mAs, and type of machine (conventional or digital); study the geographical distribution of radiological X-ray machines, and establish DRLs in Peru. Two radiological procedures were considered, general X-rays (fixed and mobile) and intraoral X-rays (fixed, mobile, and portable). An Unfors RaySafe Xi detector (Unfors RaySafe AB, Billdal, Sweden) was used; air kerma was measured at a source to image distance that varied depending on the procedure, and the entrance skin dose was subsequently calculated using the Arcal XLIX formalism. The data were collected over a period of three years (2015-2017). Only results from the last evaluation during this period were taken into consideration for each X-ray machine. DRLs were calculated at 0.21 mSv, and 0.25 mSv for posterior-anterior chest examinations in conventional and digital machines, respectively; 4.39 mSv and 6.01 mSv for conventional and digital antero-posterior lumbar spine examinations, respectively; and at 4.21 mSv for the dental intraoral procedure. The largest amount of X-ray machines is concentrated in the city of Lima. These results reflect the standard of practice in Peru.

摘要

自1996年以来,许多国家都制定了X射线检查的诊断参考水平(DRLs)。在秘鲁,由于DRLs研究有限,文献数据被用作指导方针。本研究的目的是分析放射技术中使用的参数和变量,如千伏(kV)、毫安秒(mAs)和机器类型(传统或数字);研究放射X射线机的地理分布,并在秘鲁建立DRLs。考虑了两种放射程序,普通X射线(固定和移动)和口腔内X射线(固定、移动和便携式)。使用了一台Unfors RaySafe Xi探测器(瑞典比尔达尔的Unfors RaySafe AB公司);在源到图像的距离处测量空气比释动能,该距离根据程序而变化,随后使用Arcal XLIX形式计算皮肤入口剂量。数据收集历时三年(2015 - 2017年)。对于每台X射线机,仅考虑该期间最后一次评估的结果。传统和数字机器在后前位胸部检查中的DRLs分别计算为0.21 mSv和0.25 mSv;传统和数字前后位腰椎检查的DRLs分别为4.39 mSv和6.01 mSv;口腔内牙科程序的DRLs为4.21 mSv。数量最多的X射线机集中在利马市。这些结果反映了秘鲁的实践标准。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/3495/8571906/fac716b02d62/cureus-0013-00000018566-i01.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验