Çekmece Nuclear Research and Training Center (ÇNAEM), Istanbul, Turkey.
Istanbul Technical University (ITU), Institute of Energy, Ayazaga Kampusu, Istanbul, Turkey.
Radiat Prot Dosimetry. 2018 Dec 1;182(3):377-385. doi: 10.1093/rpd/ncy076.
The aim of this study was to evaluate the radiation doses to patient during chest (posterior anterior/and lateral) examinations. The study was performed in three public hospitals of İstanbul province with a total of 300 adult patients. Entrance surface dose (ESD) measurements were conducted on computed radiography, digital radiography and screen film system. ESD was estimated by using International Atomic Energy Agency (IAEA) model and Davies model which are the common indirect models. Results were compared with diagnostic reference levels from the European Commission, IAEA and National Radiological Protection Board. Although the results are compatible with the international diagnostic reference levels, they present variations between the hospitals. Dose variations for the same type of X-ray examination support the idea that further optimization is possible.
本研究的目的是评估胸部(前后位/和侧位)检查期间患者所接受的辐射剂量。该研究在伊斯坦布尔省的三家公立医院进行,共有300名成年患者。在计算机X线摄影、数字X线摄影和屏-片系统上进行了入射表面剂量(ESD)测量。使用国际原子能机构(IAEA)模型和戴维斯模型这两种常见的间接模型来估算ESD。将结果与欧盟委员会、国际原子能机构和国家放射防护委员会的诊断参考水平进行了比较。尽管结果与国际诊断参考水平相符,但各医院之间仍存在差异。相同类型X线检查的剂量差异表明进一步优化是可行的。