使用动态导航将锥形或直形种植体植入新鲜拔牙窝中的准确性和初期稳定性:一项随机对照临床试验。

Accuracy and primary stability of tapered or straight implants placed into fresh extraction socket using dynamic navigation: a randomized controlled clinical trial.

机构信息

Department of Oral and Maxillo-Facial Implantology, Shanghai Ninth People's Hospital, College of Stomatology, Shanghai Jiao Tong University School of Medicine, National Clinical Research Center for Oral Diseases, Shanghai Laboratory of Stomatology & Shanghai Research Institute of Stomatology, Shanghai, China.

出版信息

Clin Oral Investig. 2022 Mar;26(3):2733-2741. doi: 10.1007/s00784-021-04247-2. Epub 2021 Nov 19.

Abstract

OBJECTIVES

To compare the accuracy and primary stability of tapered and straight implants undergoing immediate implant placement with dynamic navigation.

MATERIALS AND METHODS

Patients with compromised anterior teeth in maxilla were recruited and allocated randomly into (1) tapered implant group (TI group) and (2) straight implant group (SI group). Implants were inserted into fresh sockets with dynamic navigation. Three-dimensional platform deviation, apex deviation, angular deviation, insertion torque value (ITV) and implant stability quotient (ISQ) were recorded.

RESULTS

Twenty patients with 20 implants were included. The overall platform, apex, and angular deviation were 0.87 ± 0.35 mm, 0.81 ± 0.34 mm, and 2.40 ± 1.31°, respectively. The accuracy was 0.86 ± 0.26 mm, 0.76 ± 0.33 mm, and 2.49 ± 1.54° for TI, and 0.89 ± 0.44 mm, 0.88 ± 0.36 mm, and 2.31 ± 1.01° for SI, with no significant difference (p = 0.85, 0.45, 0.76). Sagittal root position classification (SRP) class I may obtain greater error in numerical values in straight implants (0.97 ± 0.47 mm vs. 0.6 ± 0.16 mm, 0.92 ± 0.36 mm vs. 0.73 ± 0.36 mm, 2.48 ± 1.19° vs. 1.71 ± 0.14°). The overall ISQ was 60.74. ISQ was 60.48 for TI and 60.96 for SI, with no significant difference. Acceptable ITV (> 15 Ncm) was achieved in most of the included patients (SI 7/10, TI 9/10).

CONCLUSIONS

High accuracy and primary stability of immediate implant placement could be achieved both in tapered and straight implants with dynamic navigation systems.

CLINICAL RELEVANCE

Tapered and straight implants did not reach a consensus on which was better in immediate implant regarding to accuracy and primary stability. Our study demonstrated implant macrodesign did not affect accuracy and primary stability in immediate implant using dynamic navigation.

摘要

目的

比较使用动态导航进行即刻种植时锥形和直形种植体的准确性和初期稳定性。

材料和方法

招募上颌前牙缺损的患者,并随机分为(1)锥形种植体组(TI 组)和(2)直形种植体组(SI 组)。使用动态导航将种植体插入新鲜牙槽窝中。记录三维平台偏差、根尖偏差、角度偏差、植入扭矩值(ITV)和种植体稳定性指数(ISQ)。

结果

共纳入 20 例患者的 20 枚种植体。总的平台、根尖和角度偏差分别为 0.87±0.35mm、0.81±0.34mm 和 2.40±1.31°。TI 的准确性为 0.86±0.26mm、0.76±0.33mm 和 2.49±1.54°,SI 的准确性为 0.89±0.44mm、0.88±0.36mm 和 2.31±1.01°,差异无统计学意义(p=0.85、0.45、0.76)。矢状根位置分类(SRP)I 级在直形种植体中可能会获得更大的数值误差(0.97±0.47mm 比 0.6±0.16mm、0.92±0.36mm 比 0.73±0.36mm、2.48±1.19°比 1.71±0.14°)。总体 ISQ 为 60.74。TI 的 ISQ 为 60.48,SI 的 ISQ 为 60.96,差异无统计学意义。大多数纳入的患者均获得了可接受的 ITV(>15Ncm)(SI 7/10,TI 9/10)。

结论

使用动态导航系统,锥形和直形种植体均可实现即刻种植的高精度和初期稳定性。

临床意义

锥形和直形种植体在即刻种植的准确性和初期稳定性方面尚未达成共识。本研究表明,在使用动态导航的即刻种植中,种植体的宏观设计不会影响准确性和初期稳定性。