Mura Marco, Martin Wendy, Williams K Keven, Abulon Dina Joy K
Department of Translational Medicine, University of Ferrara, Ferrara, Italy.
Department of Ophthalmology, University of Illinois, Chicago, IL, USA.
Clin Ophthalmol. 2021 Nov 19;15:4499-4505. doi: 10.2147/OPTH.S327570. eCollection 2021.
Novel 3D digital display systems, such as NGENUITY 3D digitally assisted visualization system (DAVS), can provide enhanced illumination, depth of field, and digital filtering. This study compared vitreous removal using NGENUITY 3D DAVS with a standard surgical microscope.
This was an in vivo, 2-arm laboratory study in 15 Yorkshire pig eyes. The LuxOR LX3 microscope with NGENUITY 3D DAVS (arm 1) and the LuxOR LX3 microscope alone (arm 2) were used with 5× optical magnification and Oculus BIOM 200 mm optics. Standard core and peripheral vitrectomy without scleral depression was performed using the CONSTELLATION Vision System. Residual vitreous weight was assessed in enucleated eyes by a masked observer. Axial length and vitreous weight of contralateral eyes were compared from an additional 14 Yorkshire pigs to confirm that eyes from a single animal were essentially identical.
After vitrectomy, mean ± SD residual vitreous was significantly smaller with NGENUITY versus standard microscope (0.143 ± 0.146 versus 0.580 ± 0.269 g, respectively; < 0.0001). Based on a mean initial vitreous weight of 2.5 g, as determined by assessment in contralateral eyes from an additional 14 Yorkshire pigs, the mean percentage vitreous removal was 94% ± 6% versus 77% ± 11%, respectively. Further, vitreous weight and axial length for contralateral eyes from any single animal in these additional 14 animals were essentially identical, as mean differences were 0.046 ± 0.035 g and 0.11 ± 0.08 mm, respectively.
Vitrectomy with NGENUITY 3D DAVS resulted in significantly less residual vitreous in pig eyes compared with standard microscopy. NGENUITY may improve vitreous removal during vitreoretinal surgery by enhancing visualization.
新型3D数字显示系统,如NGENUITY 3D数字辅助可视化系统(DAVS),可提供增强的照明、景深和数字滤波功能。本研究比较了使用NGENUITY 3D DAVS与标准手术显微镜进行玻璃体切除的效果。
这是一项在15只约克郡猪眼中进行的体内双臂实验室研究。使用配备NGENUITY 3D DAVS的LuxOR LX3显微镜(第1组)和单独的LuxOR LX3显微镜(第2组),光学放大倍数为5倍,采用Oculus BIOM 200 mm光学系统。使用CONSTELLATION视觉系统在不进行巩膜压迫的情况下进行标准的核心和周边玻璃体切除术。由一位不知情的观察者对摘除眼球中的残余玻璃体重量进行评估。比较另外14只约克郡猪对侧眼的眼轴长度和玻璃体重量,以确认来自同一只动物的眼睛基本相同。
玻璃体切除术后,与标准显微镜相比,使用NGENUITY时平均±标准差的残余玻璃体明显更少(分别为0.143±0.146克和0.580±0.269克;<0.0001)。根据另外14只约克郡猪对侧眼评估确定的平均初始玻璃体重量为2.5克,平均玻璃体切除百分比分别为94%±6%和77%±11%。此外,在这另外14只动物中,来自任何单只动物的对侧眼的玻璃体重量和眼轴长度基本相同,平均差异分别为0.046±0.035克和0.11±0.08毫米。
与标准显微镜相比,使用NGENUITY 3D DAVS进行玻璃体切除术在猪眼中产生的残余玻璃体明显更少。NGENUITY可能通过增强可视化效果来改善玻璃体视网膜手术中的玻璃体切除。