Suppr超能文献

A positron camera using position-sensitive detectors: PENN-PET.

作者信息

Muehllehner G, Karp J S

出版信息

J Nucl Med. 1986 Jan;27(1):90-8.

PMID:3484522
Abstract

A single-slice positron camera has been developed with good spatial resolution and high count rate capability. The camera uses a hexagonal arrangement of six position-sensitive NaI(Tl) detectors. The count rate capability of NaI(Tl) was extended to 800k cps through the use of pulse shortening. In order to keep the detectors stationary, an iterative reconstruction algorithm was modified which ignores the missing data in the gaps between the six detectors and gives artifact-free images. The spatial resolution, as determined from the image of point sources in air, is 6.5 mm full width at half maximum. We have also imaged a brain phantom and dog hearts.