Rowell Shannon L, Reyes Christopher R, Hopper Robert H, Engh Charles A, Muratoglu Orhun K
Harris Orthopaedics Laboratory at Massachusetts General Hospital, Boston, Massachusetts, USA.
Anderson Orthopaedic Research Institute, Alexandria, Virginia, USA.
J Biomed Mater Res B Appl Biomater. 2022 May;110(5):1113-1119. doi: 10.1002/jbm.b.34985. Epub 2021 Dec 11.
Crosslinking substantially reduces the wear of ultra-high molecular weight polyethylene (UHMWPE) used in total hip arthroplasty (THA) but some reports have indicated that first generation liners manufactured without antioxidants may be vulnerable to in vivo oxidation. This study evaluated maximum oxidation using Fourier transform infrared spectroscopy per ASTM F2102-06 and linear head penetration using a coordinate measuring machine among 66 revision-retrieved THA components with in vivo durations ranging from 0.02 to 24.6 years. These included 30 liners crosslinked with 5 Mrad of gamma radiation and then melted, 13 non-crosslinked, never-irradiated liners sterilized with gas plasma and 23 non-crosslinked, never-irradiated liners sterilized with ethylene oxide. All liners were vacuum-sealed and stored at -20°C prior to analysis with the exception of three retrievals of each material type that were stored in air for 9.9 to 21.5 years. All 57 vacuum-sealed and frozen retrievals demonstrated good oxidative stability with maximum oxidation indices (OIs) less than 1.0 and 75% (43/57) of these liners had maximum OIs less than 0.1. Linear penetration measurements were lower in the crosslinked liners compared to non-crosslinked retrievals. Although instances of oxidation and embrittlement were found after ex vivo storage in air among liners that did not have free radicals at the time of implantation, in vivo oxidation does not appear to be a clinical concern through the first decade of service for crosslinked liners and at up to 25 years after surgery for non-crosslinked liners.
交联显著降低了全髋关节置换术(THA)中使用的超高分子量聚乙烯(UHMWPE)的磨损,但一些报告表明,第一代未添加抗氧化剂制造的髋臼衬垫可能易受体内氧化影响。本研究使用傅里叶变换红外光谱法(按照ASTM F2102 - 06标准)评估了66个翻修取出的THA组件的最大氧化程度,并使用坐标测量机测量了线性头部穿透情况,这些组件的体内使用时间为0.02至24.6年。其中包括30个经5兆拉德伽马辐射交联后再熔融的髋臼衬垫、13个未交联且从未辐照过的经气体等离子体灭菌的髋臼衬垫以及23个未交联且从未辐照过的经环氧乙烷灭菌的髋臼衬垫。除每种材料类型各有3个样本在空气中储存9.9至21.5年外,所有髋臼衬垫在分析前均真空密封并储存在-20°C。所有57个真空密封并冷冻保存的样本均表现出良好的氧化稳定性,最大氧化指数(OI)小于1.0,其中75%(43/57)的髋臼衬垫最大OI小于0.1。与未交联的取出样本相比,交联髋臼衬垫的线性穿透测量值更低。尽管在植入时没有自由基的髋臼衬垫在体外空气中储存后出现了氧化和脆化情况,但对于交联髋臼衬垫,在服务的第一个十年以及对于未交联髋臼衬垫在术后长达25年的时间里,体内氧化似乎并非临床问题。