超短回波时间磁共振血管成像结合减影法评估血流导向装置治疗颅内动脉瘤。

Ultra-short Echo-time MR Angiography Combined with a Subtraction Method to Assess Intracranial Aneurysms Treated with a Flow-diverter Device.

机构信息

Central Division of Radiology, Jichi Medical University Saitama Medical Center.

Department of Radiology, National Defense Medical College.

出版信息

Magn Reson Med Sci. 2023 Jan 1;22(1):117-125. doi: 10.2463/mrms.tn.2021-0106. Epub 2021 Dec 11.

Abstract

A flow-diverter (FD) device is a well-established tool for the treatment of unruptured intracranial aneurysms. Time-of-flight (TOF) MR angiography (MRA) is widely used for postoperative assessment after the treatment with FD; however, it cannot fully visualize intra-aneurysmal and intrastent flow signals due to the magnetic susceptibility from the FD. Recently, the utility of MRA with ultra-short TE (UTE) sequence and arterial spin labeling technique in assessing the therapeutic efficacy of intracranial aneurysms treated with metallic devices has been reported, but long image acquisition time is one of the drawbacks of this method. Herein, we introduce a novel UTE MRA using the subtraction method that enables the reduction in susceptibility artifacts with a short image acquisition time.

摘要

血流导向装置(FD)是一种成熟的治疗未破裂颅内动脉瘤的工具。在 FD 治疗后,磁共振血管造影(MRA)中的时间飞越(TOF)法被广泛用于术后评估;然而,由于 FD 的磁敏感性,它不能完全显示瘤内和支架内的血流信号。最近,报道了使用超短回波时间(UTE)序列和动脉自旋标记技术的 MRA 在评估金属器械治疗颅内动脉瘤的治疗效果中的应用,但是该方法的一个缺点是采集时间较长。在此,我们介绍了一种新的 UTE MRA 减影法,该方法能够在短时间内减少磁敏感性伪影。