Department of Surgical Sciences/Section of Orthopaedics, Uppsala University, Uppsala, Sweden.
Acta Orthop. 2022 Jan 3;93:206-211. doi: 10.2340/17453674.2021.1080.
Background and purpose - We previously described a decrease in bone mineral density (BMD) in the calcar region 2 years after insertion of the collum femoris-preserving (CFP) stem, but the implants were stable. Now we have examined the long-term changes in periprosthetic BMD and stability of the CFP stem. Patients and methods - We conducted a minimum 8-year follow-up of 21 patients from our original investigation. We examined periprosthetic BMD by dual-energy X-ray absorptiometry (DEXA) and implant stability by radiostereometric analysis (RSA). Results - Between 2 and 8 years 1 stem was revised due to aseptic loosening. Between 2 and 8 years we found a 14% (95% confidence interval [CI] 9-19) reduction in BMD in Gruen zone 6 and 17% (CI 6-28) in Gruen zone 7. From baseline the reduction in BMD was 30% (CI 23-36) in Gruen zone 6, 39% (CI 31-47) in Gruen zone 7, and 19% (CI 14-23) in Gruen zone 2. Between 2 and 8 years, RSA (n = 17) showed a mean translation along the stem axis of 0.02mm (CI -0.02 to 0.06) and a mean rotation around the stem axis of 0.08° (CI -0.26 to 0.41). From baseline mean subsidence was 0.07 mm (CI -0.16 to 0.03) and mean rotation around the stem axis was 0.23° (CI -0.23 to 0.68) at 8 years. Interpretation - There was continuous loss of proximomedial BMD at 8 years while the CFP stem remained stable. Proximal periprosthetic bone loss cannot be prevented by this stem.
背景与目的-我们之前描述了在股骨颈保留(CFP)柄植入后 2 年内,骨钙区域的骨密度(BMD)下降,但植入物是稳定的。现在我们已经检查了 CFP 柄的长期周围 BMD 变化和稳定性。
患者和方法-我们对我们原始研究中的 21 名患者进行了至少 8 年的随访。我们通过双能 X 射线吸收法(DEXA)检查周围 BMD,通过放射立体分析(RSA)检查植入物稳定性。
结果-在 2 至 8 年内,1 个柄因无菌性松动而翻修。在 2 至 8 年内,我们发现 Gruen 区 6 的 BMD 减少了 14%(95%置信区间[CI]9-19),Gruen 区 7 的 BMD 减少了 17%(CI 6-28)。从基线开始,Gruen 区 6 的 BMD 减少了 30%(CI 23-36),Gruen 区 7 的 BMD 减少了 39%(CI 31-47),Gruen 区 2 的 BMD 减少了 19%(CI 14-23)。在 2 至 8 年内,RSA(n = 17)显示沿柄轴的平均平移为 0.02mm(CI-0.02 至 0.06),围绕柄轴的平均旋转为 0.08°(CI-0.26 至 0.41)。从基线开始,平均下沉为 0.07mm(CI-0.16 至 0.03),8 年后,围绕柄轴的平均旋转为 0.23°(CI-0.23 至 0.68)。
解释-在 8 年内,近端内侧 BMD 持续丢失,而 CFP 柄保持稳定。这种柄不能防止近端周围骨质丢失。