Evaluation of radiation dose, focal spot, and automatic exposure of newer film-screen mammography units.

作者信息

Kimme-Smith C, Bassett L W, Gold R H

机构信息

Department of Radiological Sciences, University of California, Los Angeles, School of Medicine 90024.

出版信息

AJR Am J Roentgenol. 1987 Nov;149(5):913-7. doi: 10.2214/ajr.149.5.913.

Abstract

Fifteen recently introduced mammographic units representing eight different models and manufacturers were evaluated to see whether the equipment performance equaled claims made by the manufacturers. Focal-spot size was measured by pinhole and slit methods. Eight automatic exposure control systems on five models were tested by measuring the optical density produced by imaging at 28 and 30 kVp, with and without a grid and with test objects of varying thicknesses of BR12 epoxy material. Comparable mean glandular dose for a 4-cm-thick BR12 test object was measured for 13 of the units under manual or automatic exposures. Nominal focal-spot size was larger than stated for six of the 15 large focal spots and for seven of the 12 microfocal spots. Comparable doses under automatic exposure control varied more than 100% between units from different manufacturers, while the doses of units from the same manufacturer varied less than 20%. Automatic exposure control systems were operational in eight of the 15 units, but four of these eight were found to be out of calibration. There are considerable performance variations that need to be evaluated before acceptance and use of new mammographic equipment.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验