Heart Vessels. 2022 Jul;37(7):1115-1124. doi: 10.1007/s00380-021-02019-2. Epub 2022 Jan 10.
To evaluate the feasibility of spectral imaging with dual-layer spectral detector computed tomography (CT) for the diagnosis of acute coronary syndrome. We identified 30 consecutive patients who underwent cardiac CT using dual-layer spectral detector CT and were diagnosed with acute ischemic syndrome by an invasive coronary angiography. We reconstructed 120 kVp images and generated virtual monochromatic images (VMIs; 40-200 keV in 10 keV increments), iodine concentration maps, and effective atomic number (Z) maps. We calculated the contrast and contrast-to-noise ratio (CNR) between myocardial normal and hypo-perfusion and chose the VMIs with the best CNR for quantitative analysis. We compared the image noise, contrast, and CNR of 120 kVp images and the best VMIs, CT value, iodine concentration, and effective Z between myocardial normal and hypo-perfusion with the paired t test. As the X-ray energy decreased, venous attenuation, contrast, and CNR gradually increased. The 40 keV image yielded the best CNR. The contrast and CNR between myocardial normal and hypo-perfusion were significantly higher in 40 keV images than those in 120 kVp images. The iodine concentration and the effective Z were significantly higher in normal myocardium than those in hypo-perfused myocardium. Spectral imaging with dual-layer spectral detector CT is a feasible technique to detect the hypo-perfused area of acute ischemic syndrome.
评估双层光谱探测器 CT 能谱成像用于诊断急性冠状动脉综合征的可行性。我们共纳入 30 例连续因急性缺血综合征而行有创冠状动脉造影术的患者,这些患者均接受过双层光谱探测器 CT 心脏 CT 检查。我们重建了 120kVp 图像,并生成了虚拟单能量图像(VMIs;40-200keV,10keV 递增)、碘浓度图和有效原子序数(Z)图。我们计算了心肌正常和低灌注区之间的对比和对比噪声比(CNR),并选择 CNR 最佳的 VMIs 进行定量分析。我们使用配对 t 检验比较了 120kVp 图像与最佳 VMIs 的图像噪声、对比度和 CNR、心肌正常和低灌注区的 CT 值、碘浓度和有效 Z。随着 X 射线能量的降低,静脉衰减、对比度和 CNR 逐渐增加,40keV 图像的 CNR 最佳。40keV 图像的心肌正常和低灌注区之间的对比度和 CNR 明显高于 120kVp 图像。正常心肌的碘浓度和有效 Z 明显高于低灌注心肌。双层光谱探测器 CT 能谱成像技术是一种可行的检测急性缺血综合征低灌注区的技术。