To compare intraocular pressure (IOP) measurement by ORA-IOPcc and Corvis-bIOP after femtosecond laser-assisted LASIK (FS-LASIK).
In this prospective cohort study, 56 eyes from 56 consecutive patients scheduled for FS-LASIK were enrolled. All patients had IOP measurement with ORA and Corvis ST by two blinded independent expert examiners. IOP examinations were conducted between 8 and 11 A.M. Data were collected at baseline and 3 months after FS-LASIK.
The mean age of the participants was 29.1 ± 6.3 years, and 42 (75%) were female. The average of central corneal thickness (CCT) decreased from 537 ± 23 µm at baseline to 458 ± 31 µm after FS-LASIK. The mean postoperative change of IOP was 0.0 ± 2.1 for bIOP and -2.5 ± 3.2 mmHg for IOPcc. The corresponding 95% limits of agreement (LoA) was -4.1 to 4.1 mmHg and -3.8 to 8.8 mmHg, respectively. Both methods showed no significant correlation between ∆IOP and ∆CCT. The 95% LoA between bIOP and IOPcc after FS-LASIK was -4.8 to 9.1 mmHg.
Compared to the ORA-IOPcc, the Corvis-bIOP showed less variation after FS-LASIK and might be a more appropriate choice for measuring IOP in this condition. The agreement of bIOP vs. IOPcc after FS-LASIK is below the clinically acceptable level, and the two methods could not be regarded as interchangeable.