Gärtner Lutz, Lenarz Thomas, Ivanauskaite Jurgita, Büchner Andreas
Department of Otolaryngology, Hannover Medical School, Hannover, Germany.
Cochlear Implants Int. 2022 May;23(3):165-172. doi: 10.1080/14670100.2022.2026025. Epub 2022 Jan 22.
To describe the influence of stimulus parameters on aberrant facial nerve stimulation in cochlear implant users.
Retrospective case series (three ears - two patients). Cochlear implant patients with severe facial nerve stimulation and degraded speech comprehension who underwent re-implantation with Neuro Zti EVO (Oticon Medical).
In all three ears, side effects evoked by aberrant facial nerve stimulation could be resolved and thereby speech comprehension was improved.
Aberrant facial nerve stimulation in cochlear implant patients was successfully resolved by re-implantation. We hypothesize that one or more of the specific stimulus paramet ers are responsible for this success: combined common ground and monopolar stimulation mode, asymmetric pulse waveform with active anodic leading first phase followed by slow capacitive discharge, and pulse width modulation for loudness coding.